Discrepancies in the measurement of isovolumic relaxation time: a study comparing M mode and Doppler echocardiography.

نویسندگان

  • C H Lee
  • F Vancheri
  • M S Josen
  • D G Gibson
چکیده

Mitral valve cusp separation on M mode echogram, the mitral valve opening artefact, and the onset of forward transmitral flow recorded by Doppler echocardiography have all been taken to mark the end of isovolumic relaxation, while its onset has been taken either as the aortic closure sound (A2) recorded phonocardiographically or the aortic closure artefact determined by Doppler technique. Possible differences in the measurement of the isovolumic relaxation time were studied when these landmarks were used in 44 healthy people, 14 patients with mitral stenosis, 21 patients with left ventricular hypertrophy, and 24 patients with dilated cardiomyopathy by recording M mode echograms of the mitral valve, and pulsed and continuous wave Doppler spectra of transmitral flow, with simultaneous electrocardiograms and phonocardiograms. A2 was effectively synchronous with the aortic artefact. However, when the onset of Doppler flow was regarded as the end of isovolumic relaxation, the interval was significantly longer than when mitral cusp separation on M mode echograms was used: by 25 (10) ms in healthy individuals, by 25 (15) ms in patients with left ventricular hypertrophy, and by 50 (35) ms in patients with dilated cardiomyopathy. In patients with mitral stenosis the interval was only 5 (5) ms longer. The mitral valve opening artefact consistently followed the onset of flow and corresponded much more closely to the E point on the M mode echogram. This shows that it occurred during the rapid filling period and well beyond isovolumic relaxation by any definition. Thus isovolumic relaxation time measured from A2 to the onset of transmitral flow or the mitral valve opening artefact differs from that derived from A2 to mitral valve cusp separation. These intervals cannot be used interchangeably to measure "isovolumic relaxation time".

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Doppler-Derived Myocardial Performance Index in Healthy Children in Shiraz

Background: Assessment of myocardial function is essential in heart disease, but in regard to systolic and diastolic functions such evaluation has limitation. Ejection fraction is difficult to assess in abnormally-shaped ventricles, and diastolic inflow velocity pattern may be fused because of tachycardia. Objective: A myocardial performance index (MPI) or Tei index has been developed for adult...

متن کامل

Echocardiographic Measurement of Systolic Time Intervals in Healthy Great Dane Dogs

Objective- to determine the range of systolic times intervals using Pulsed-wave Doppler and m-mode echocardiography in normal, healthy Great Dane dogs. Design- Descriptive study. Animals- Thirty healthy Great Dane dogs. Procedures- This study examined 30 clinically healthy Great Dane dogs by the twodimensional, M-mode and Pulsed-wave Doppler echocardiography. The dogs were awake and unsedated. ...

متن کامل

The influence of Asthma on Right Ventricular Systolic and Diastolic Function in Children according to Tissue Doppler Echocardiography

Introduction:Asthma is the most common chronic inflammatory condition affecting the lower airways among children. Asthma may influence right ventricular (RV) function and patients may develop right ventricular hypertrophy, pulmonary hypertension and cor pulmonale. The main objective of this study is to determine RV dysfunction in asymptomatic asthmatic children as detected by ti...

متن کامل

Comparison between M-Mode and Doppler echocardiography in diagnosis

Introduction: Patent ductus arteriosus (PDA) is the persistence of a normal fetal structure between left pulmonary artery and descending aorta. Persistence of this fetal structure beyond 10 days of life is considered abnormal. In this study we compared M-Mode and Doppler echocardiography in diagnosis of PDA in children at Aliasghar Hospital in Zahedan and Children hospital in Bandar Abbas for...

متن کامل

Relation of left ventricular isovolumic relaxation time and incoordination to transmitral Doppler filling patterns.

OBJECTIVE To investigate factors during isovolumic relaxation that determine Doppler filling patterns in patients with left ventricular disease, and thus to identify the underlying mechanisms. DESIGN 85 patients (50 ischaemic heart disease, 35 left ventricular hypertrophy due to aortic stenosis) and 26 controls were studied with Doppler and M mode echocardiography and phonocardiography. 16 pa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British heart journal

دوره 64 3  شماره 

صفحات  -