The Durom large diameter head acetabular component: early results with a large-diameter metal-on-metal bearing.

نویسندگان

  • C Berton
  • J Girard
  • N Krantz
  • H Migaud
چکیده

Implantation of a large-diameter femoral head prosthesis with a metal-on-metal bearing surface reduces the risk of dislocation, increases the range of movement, minimises the risk of impingement and, in theory, results in little wear. Between February 2004 and March 2007 we implanted 100 consecutive total hip replacements with a metal-on-metal bearing and a large femoral head into 92 patients. There were 51 men and 41 women with a mean age of 50 years (18 to 70) at the time of surgery. Outcome was assessed using the Western Ontario McMaster University osteoarthritis index and the Harris hip score as well as the Devane activity score. These all improved significantly (p < 0.0001). At the last follow-up there were no cases of dislocation, no impingement, a good range of movement and no osteolysis, but seven revisions, two for infection and five for aseptic loosening. The probability of groin pain increased if the other acetabular component inclination exceeded 50 degrees (p = 0.0007). At 4.8 years of follow-up, the projected survival of the Durom acetabular component, with revision for any reason, was 92.4% (sd 2.8) (95% confidence interval 89.6 to 95.2). The design of the component made it difficult both to orientate and seat, which when combined with a poor porous coating, produced unpredictable fixation and a low survival at five years.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A retrieval analysis of explanted Durom metal-on-metal hip arthroplasties.

Given the recent reports of high failure rates, an improved understanding of the mechanism of failure of large diameter metal-on-metal hip arthroplasties is essential. We present clinical data and tribological analysis of a consecutive series of 74 failed large diameter metal-on-metal hips, comparing the Durom (Zimmer) with the Birmingham hip resurfacing (BHR) (Smith and Nephew). We retrospecti...

متن کامل

Large-head metal-on-metal total hip arthroplasty using the Durom acetabular component at minimum 1-year interval.

Large-head metal-on-metal total hip arthroplasty represents novel technology, and outcome data are lacking. We prospectively compared the performance of a nonmodular metal-on-metal acetabular component (Durom; Zimmer, Warsaw, IN) with a modular titanium component (Trilogy, Zimmer). All Durom components placed at our institution with minimum 1-year follow up (n = 63) were compared with an age- a...

متن کامل

Deformation of the Durom Acetabular Component and Its Impact on Tribology in a Cadaveric Model—A Simulator Study

BACKGROUND Recent studies have shown that the acetabular component frequently becomes deformed during press-fit insertion. The aim of this study was to explore the deformation of the Durom cup after implantation and to clarify the impact of deformation on wear and ion release of the Durom large head metal-on-metal (MOM) total hips in simulators. METHODS Six Durom cups impacted into reamed ace...

متن کامل

Taper junction failure in large-diameter metal-on-metal bearings

OBJECTIVES An ongoing prospective study to investigate failing metal-on-metal hip prostheses was commenced at our centre in 2008. We report on the results of the analysis of the first consecutive 126 failed mated total hip prostheses from a single manufacturer. METHODS Analysis was carried out using highly accurate coordinate measuring to calculate volumetric and linear rates of the articular...

متن کامل

Metal ion release from bearing wear and corrosion with 28 mm and large-diameter metal-on-metal bearing articulations: a follow-up study.

We have updated our previous randomised controlled trial comparing release of chromium (Cr) and cobalt (Co) ions and included levels of titanium (Ti) ions. We have compared the findings from 28 mm metal-on-metal total hip replacement, performed using titanium CLS/Spotorno femoral components and titanium AlloFit acetabular components with Metasul bearings, with Durom hip resurfacing using a Meta...