The Methodology for Evaluating Half-value Layer and Its Influence on the Diagnostic Radiology*
نویسندگان
چکیده
OBJECTIVE: The critical analysis of the methodology for evaluating the x-ray beam half-value layer was based on technical standards for radiological protection and quality control in medical and dental diagnoses. MATERIALS AND METHODS: Exposure parameters, experimental set-up, scattering materials, instrumentation, size of radiation field and methodology of data analysis have been taken into consideration for measurements of half-value layer. RESULTS: The half-value layer, in good geometry, measured 2.44 ± 0.02 mmAl. A maximum deviation of 4.1% was found under other geometry conditions. A 98.8% deviation was observed with a different ionization chamber, and scattering material, and 29.5% with another type of detector. CONCLUSION: Results showed that no relevant variation is observed in the half-value layer for different radiation field sizes. On the other hand, the influence of the radiation scattering on the half-value layer overestimation was clearly observed, as well as the half-value layer reduction in the presence of lead used as scattering material. The practical methodology adopted for half-value layer evaluation has shown to be quite reliable, besides evidencing remarkable discrepancies resulting from the adoption of inappropriate methodologies, emphasizing the necessity of establishing a standard procedure for measuring the half-value layer.
منابع مشابه
Investigating the effect of voltage fluctuations on the beam quality of X-rays in the range energy of diagnostic radiology
The purpose of this study was to investigate the effect of voltage ripple produced by X-ray machines on the quality of X-ray rays. For this purpose, the energy spectrum from the X-ray tube manufacturer should be simulated in the range of diagnostic radiography and verified by experimental results. The statistical analysis of the results showed that there is no significant difference between sim...
متن کاملInfluence of the Different Primary Cancers and Different Types of Bone Metastasis on the Lesion-based Artificial Neural Network Value Calculated by a Computer-aided Diagnostic System,BONENAVI, on Bone Scintigraphy Images
Objective(s): BONENAVI, a computer-aided diagnostic system, is used in bone scintigraphy. This system provides the artificial neural network (ANN) and bone scan index (BSI) values. ANN is associated with the possibility of bone metastasis, while BSI is related to the amount of bone metastasis. The degree of uptake on bone scintigraphy can be affected by the type of bone metastasis. Therefore, t...
متن کاملQuality Control Assessment of Radiology Devices in Kerman Province, Iran
Introduction Application of quality control (QC) programs at diagnostic radiology departments is of great significance for optimization of image quality and reduction of patient dose. The main objective of this study was to perform QC tests on stationary radiographic X-ray machines, installed in 14 hospitals of Kerman province, Iran. Materials and Methods In this cross-sectional study, QC tests...
متن کاملInfluence of Rigidity, Irregularity and Initial Stress on Shear Waves Propagation in Multilayered Media
The propagation of shear waves in an anisotropic fluid saturated porous layer over a prestressed semi-infinite homogeneous elastic half-space lying under an elastic homogeneous layer with irregularity present at the interface with rigid boundary has been studied. The rectangular irregularity has been taken in the half-space. The dispersion equation for shear waves is derived by using the pertur...
متن کاملQuality Control Assessment of Conventional Radiology Devices in Iran
Introduction Quality control (QC) techniques are used in monitoring and maintenance of the components of an x-ray system. QC of radiology devices plays a significant role in reduction of medication dose and optimization of image quality. This study aimed to conduct QC tests on randomly selected radiology devices, installed in diagnostic imaging departments of Iran. Materials and Methods In tota...