Acetabular cup geometry and bone-implant interference have more influence on initial periprosthetic joint space than joint loading and surgical cup insertion.

نویسندگان

  • Kevin L Ong
  • Jeffrey Lehman
  • William I Notz
  • Thomas J Santner
  • Donald L Bartel
چکیده

Environmental variations in patient-dependent and surgical factors were modeled using robust optimization with a finite element acetabular cup-pelvis model. A previously developed statistical optimization scheme was used to: (1) determine the cup geometry and the optimal cup-bone interference that maximized bone-implant contact areas and minimized changes in the gap volume between the implant and bone surface during gait loading and unloading; and (2) determine the relative contributions of design, patient-dependent, and surgical factors to variations in bone-implant contact areas and a change in gap volume. The statistical analyses indicated that the design variables, namely the equatorial diameter and eccentricity, explained most of the variations in the performance measures. Further, the hemispherical designs performed better than the nonhemispherical designs. The 58 mm hemispherical cup, with 2 mm diametral interferences, minimized the change in gap volume and attained 82% and 81% of the maximum predicted total and rim contact areas, respectively. The equatorial diameter and eccentricity, not the patient-dependent and surgical factors, explained most of the variations in the performance measures. Perfect surface apposition was not attained with any of the cup designs.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A novel method to assess primary stability of press-fit acetabular cups.

Initial stability is an essential prerequisite to achieve osseointegration of press-fit acetabular cups in total hip replacements. Most in vitro methods that assess cup stability do not reproduce physiological loading conditions and use simplified acetabular models with a spherical cavity. The aim of this study was to investigate the effect of bone density and acetabular geometry on cup stabili...

متن کامل

A technique for removing an intrapelvic acetabular cup.

We describe a simple, retroperitoneal approach for the removal of acetabular components that have migrated into the pelvis. The dense fibrous tissue layer which surrounds the implant protects the iliac vessels during removal of the cup by this method.

متن کامل

Construction and validation of a finite element model of a human pelvis

Introduction The hip joint is one of the most important joints for the mobility of the human body. Degeneration of this joint often results in immobilisation and disabling pain. In the past 30 years total hip arthroplasty has proven to reduce pain and to improve quality of life. In order to guarantee long-term stability of these devices it is important to obtain a long lasting fixation of the p...

متن کامل

The Comparative Analysis of a Novel Acetabular Component Against Hemispherical Component in Case of Extensive Acetabular Bone Defects A Study Of Finite Element Analysis

The purpose of this study was to evaluate the design of a cup using finite element method and to analyze possible effects of joint loading postoperatively, and its initial mechanical stability, so as to direct its further optimization. Finite‐element (FE) models of the cup with three wings and the hemispherical cup were created to calculate the stress pattern...

متن کامل

Press-fit acetabular cup fixation: principles and testing.

Pre-clinical testing of the fixation of press-fit acetabular components of total hip prostheses relies on cadaver or synthetic bone, but the properties and geometry of bone models differ from those of physiological bone. Cup designs use varied mechanisms for initial stability in bone; therefore, using different analogues and tests is appropriate. Press-fit cup stability was tested in the follow...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of biomechanical engineering

دوره 128 2  شماره 

صفحات  -

تاریخ انتشار 2006