An evaluation of the Vita-Stat automatic blood pressure measuring device.

نویسندگان

  • B F Polk
  • B Rosner
  • R Feudo
  • M Vandenburgh
چکیده

Four Vita-Stat (VS) automatic, coin-operated, blood pressure measuring devices were evaluated for accuracy and precision. Under field conditions, 342 adults had two blood pressure measurements with both the VS device and with the Random-Zero (RZ) device. Two of the VS machines gave significantly higher systolic (SBP) and diastolic blood pressure (DBP) values, compared to the RZ values. The mean difference between these two units and the RZ device were clinically important (14.0 and 6.9 mm Hg SBP; 7.5 and 6.8 mm Hg DBP), and resulted in the misclassification of 23% of normotensives as hypertensives. We observed a significant order effect on SBP with the VS device, with a mean decrement of 4.9 mm Hg between the first and second SBP values, compared to 0.6 mm Hg with the RZ device. Even after adjusting for this bias, all four VS devices gave significantly more variable SBP readings than the RZ unit; two of the four also gave more variable DBP values. These data suggest that the VS device is neither accurate nor precise enough at the present time to be recommended for widespread use. These findings also raise questions concerning the monitoring of performance of this and similar devices in the field.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of a community-based automated blood pressure measuring device.

BACKGROUND Automated devices are widely available in the community for people to measure their blood pressure. We assessed the accuracy and reproducibility of a brand of community-based automated device against the standard mercury sphygmomanometer. METHODS Same-arm pairs of blood pressure readings were obtained with the Vita-Stat 90550 automated device, a sphygmomanometer and the Omron HEM-7...

متن کامل

Lack of alerting reactions to intermittent cuff inflations during noninvasive blood pressure monitoring.

Intermittent noninvasive blood pressure monitoring is becoming increasingly popular because of the belief that a daytime blood pressure profile can provide a better clinical evaluation of hypertension than that provided by casual blood pressure measurements. This approach has potential limitations, however, one of which is that the cuff inflations permitting blood pressure to be repeatedly meas...

متن کامل

گزارش یک مورد: ارسال فشار خون و نبض بیمار به پزشک با استفاده از سیستم Global system for Mobile Communications

 Background: Hypertension is a common chronic disease with increasing prevalence worldwide. Regular close monitoring of blood pressure is key point of hypertension management. Home monitoring of blood pressure has its own problems such as lack of early access to a physician in case of emergency. In this study we have designed a device that could send the patient’s self-measured blood pre...

متن کامل

مقایسه اندازه گیری فشارخون توسط گوشی معمولی با گوشی تلفیقی

Background and Purpose: Blood pressure (BP) is routinely evaluated intermittently during the clinical status for care and treatment. The most common method to measure BP is by using the sphygmomanometer with a stethoscope. This method may be difficult for medical students that will use BP for medical evaluation. The purpose of this study was to compare the accuracy of double stethoscope devic...

متن کامل

Comparison of Various Blood Pressure Measurement Techniques with Standard Office Based Blood Pressure Mesurement

Introduction: In order to control Hypertension, it is necessary to understand the methods of blood pressure measurement. Recent guidelines emphasize on employing appropriate blood pressure measurement techniques and white coat hypertension. Definition and methods of blood pressure measurement have changed in the American guideline based on SPRINT trial (Systolic Blood Pressure Intervention Tria...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Hypertension

دوره 2 2  شماره 

صفحات  -

تاریخ انتشار 1980