This paper compares and contrasts the use of MOSFET dosimeters and extended rem counters for neutron detection in cosmic radiation fields. These technologies have been exposed to a reference field and data are presented that demonstrate the contrasting sensitivities of these approaches. The MOSFETs are observed to be insensitive to dose levels associated with cosmic fields whilst the extended rem counters give a good response, demonstrating that copper has potential as a replacement for lead in these systems.
منابع مشابه
The effect of source shield on landmine detection
Background: Several landmine detection methods, based on nuclear techniques, have been suggested during the recent years. Neutron energy moderation, neutron-induced gamma emission, neutron and gamma attenuation, and fast neutron backscattering are nuclear-based methods used for landmine detection. The aim of this study is to use backscattered neutron for landmine detection. Materials ...
متن کاملThe safety of a landmine detection system using graphite and polyethylene moderator
Background: Several landmine detection methods, based on nuclear techniques, have been suggested up to now. Neutron-induced gamma emission, neutron and gamma attenuation, and fast neutron backscattering are the nuclear methods used for landmine detection. In this paper an optimized (safe and effective) moderating structure using an 241Am-Be neutron source for detecting landmines has been invest...
متن کاملCancer Risk Assessment due to Accidental Exposure inside Neutron Laboratories using BEIR VII Model
Introduction: Environmental and occupational human exposure from neutron source can lead to the serious biologic effects. The aim of this study is to evaluate the cancer incidence risk for various human organs at different neutron dose levels due to exposure from an Americium-241/Beryllium (Am-241/Be), a standard neutron source for calibration purposes. Material and Methods: We measured ambient...
متن کاملInvestigation of neutron flux by thermoluminescence dosimeters in the neutron-gamma mixed field
Introduction: Neutrons have many applications in various fields, such as medicine. The important application of neutron in medical science is in Boron Neutron Capture Therapy (BNCT). Because of this, determination of neutron dose and flux is critical for the health maintaining of workers and patients exposed to this beam. The neutron radioisotope sources produce gamma rays in ...
متن کاملSimulation of a Neutron Detector for Real Time Imaging Applications
Monte Carlo Method is used to simulate a double layer gadolinium-amorphous silicon thermal neutron detector. The detector fabricated in pixel array configuration has various applications including neutron imaging. According to the simulation results, a detector consisting of a gadolinium (Gd) film with thickness of 2-4 ~m, sandwiched properly with two layers of sufficiently thick (-30 ?µm) hydr...
متن کامل