Fracture at the input tube-cylinder junction of AMS 700 inflatable penile prostheses as a complication of a modified implantation technique in a series of 99 patients.

نویسندگان

  • S C Kim
  • K K Seo
  • S H Yoon
چکیده

OBJECTIVES To compare the incidence of a specific failure mode of the penile implant, fracture at the input tube-cylinder junction, with respect to two methods of managing the input tube. METHODS AMS 700 series three-piece inflatable penile prostheses were implanted in the first 26 patients using an ordinary technique in which the input tubing runs alongside the cylinder within the corpus and exits through the corporotomy (method A). In the subsequent 73 men, the input tube exited through a separate stab wound in the proximal corpus using a modification of the basic surgical technique (method C). The mean follow-up period was 136.4 months for method A and 69.0 months for method C. The incidence of fracture at the junction of the input tube and cylinder was compared according to the variables of input tube management, prosthesis type, and width of the proximal corpora. RESULTS The overall incidence of mechanical failure was 12.1%. Fractures at the input tube-cylinder junction with leaking occurred in 7 patients. The cylinders in these patients were all implanted using method C. The incidence of fracture at the junction was significantly higher (P <0.05) in men with narrow corpora (17.1%) than in the others (0%), regardless of the type of prosthesis implanted. The average functional duration of the failed prostheses was 66.1 months. CONCLUSIONS The modified surgical procedure (method C) should be avoided in patients with a narrow-width penis because of an increased likelihood of damage to the input tube-cylinder junction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Mechanical reliability of AMS hydraulic penile prostheses.

The mechanical reliability of AMS hydraulic penile prostheses implanted in 203 patients from April 1985 to June 1995 were evaluated. AMS Hydroflex prosthesis showed the highest incidence of mechanical failure (18.8%; 6/32 patients) during a mean follow-up period of 94.5 (64-117) months. Mean functioning time of the prostheses until malfunction was 50 (0-100) months. Bilateral fractures at junct...

متن کامل

Re: Comparison of a new length measurement technique for inflatable penile prosthesis implantation to standard techniques: outcomes and patient satisfaction.

INTRODUCTION Within a study evaluating the redesigned AMS 700MS inflatable penile prosthesis (IPP) (American Medical Systems, Minnetonka, MN, USA), one site used new length measurement technique (NLMT), a more aggressive dilation and measurement of the corpora cavernosa on a stretched penis, to address penile shortening. AIM To compare cylinder size and patient satisfaction, between a NLMT an...

متن کامل

Mechanism of Penile Prosthesis Induced Urethral Stricture: Treatment and Prevention

The advent of phosphodiesterase type 5 inhibitor has revolutionized the management of erectile dysfunction (ED). Since the introduction of sildenafil citrate in 1998, oral systemic therapy has become the first line therapy for men with ED. However, oral pharmacotherapy often fails in patients with severe diabetes, radical prostatectomy, Peyronie’s disease, and severe penile fibrosis. These pati...

متن کامل

Physician and patient satisfaction with the new AMS 700 momentary squeeze inflatable penile prosthesis.

INTRODUCTION A single-armed, prospective, multicenter study evaluated the redesigned American Medical Systems (AMS) 700 Momentary Squeeze (MS) pump inflatable penile prosthesis (IPP) with enhanced features for ease of implantation and patient manipulation. The device incorporates design changes to all components: pump, cylinders, rear-tip extenders, and reservoir. AIM To assess physician and ...

متن کامل

Penile implants in the treatment of organic impotence.

OBJECTIVE To evaluate the reliability and safety of penile implants in the treatment of organic impotence at the Saudi Aramco - Dhahran Health Center. METHODS A series of 108 cases of organic impotence that underwent 125 penile implantation procedures between 1988 and 1997 was reviewed. The follow-up period ranged between 6 months and 10 years. The mean age was 57.9 years (range 26-76). The p...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Urology

دوره 54 1  شماره 

صفحات  -

تاریخ انتشار 1999