Echocardiographic and phonocardiographic confirmation of suspected caged mitral valve malfunction.

نویسندگان

  • T B Berndt
  • D J Goodman
  • R L Popp
چکیده

Seven patients studied by echocardiography with and without simultaneous phonocardiography for suspected malfunction of a caged mitral valve prosthesis are presented. In case 1, with inaudible prosthetic clicks, thrombosis of the cage and immobility of the ball were suggested by echocardiographic studies and confirmed at surgery. In case 2, simultaneous echocardiographic and phonocardiographic studies demonstrated wide and variable intervals between the aortic second sound the the opening click and also "sticking" of the ball. In case 3 a thrombus prevented full motion of the ball to the apex of the cage, which was seen on the echocardiogram, while in case 4, with a thrombus within the ventricle and prosthesis, the prosthetic opening click was present intermittently and was associated with only subtle echocardiographic changes. In case 5, echocardiographic studies demonstrated abnormal rocking of the cage secondary to severe prosthetic dehiscence. In case 6, dul prosthetic clicks were to be secondary to a low cardiac-output state. In case 7, with multiple valve prostheses, simultaneous echocardiographic and phonocardiographic studies allowed identification of individual valve sounds and abnormal timing of valve opening. Based on these studies, we believe that echocardiography and simultaneous phonocardiography can yield very useful information in the evaluation of patients with suspected malfunction of a caged mitral valve prosthesis.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Impaired function of cloth-covered Starr-Edwards mitral valve prosthesis. Detection by phonocardiography.

was suspected because of phonocardiographic abnormalities. A prolonged interval from aortic closure to opening of the mitral ball suggested an impediment to movement of the ball. This was confirmed by catheterization and cinefluorography. At autopsy the impaired ball movement was found to be due to thrombus buildup on the outflow face of the mitral valve orifice. It is suggested that frequent p...

متن کامل

Effect of position on echocardiographic diagnosis of mitral valve prolapse

Echocardiography has been accepted as the diagnostic modality of choice for mitral valve prolapse (MVP). The determine the effect of position on echocardiographic diagnosis of MVP, we examined 693 consecutive patients referred to the echocardiography laboratory because of suspected MVP. First, patients were examined using the usual standard M-Mode echocardiography( left parasternal third or fou...

متن کامل

Diagnosis of prosthetic mitral valve malfunction with combined echo-phonocardiography.

Fifty-three patients were studied with combined echo-phonocardiography or phonocardiography alone following prosthetic valve replacement. In sixteen of these patients, clinical deterioration developed, and all subsequently underwent cardiac catheterization and/or surgery. Two patients came to autopsy. Included in this group of sixteen patients were five with obstructed prosthesis, six with para...

متن کامل

Mitral valve prolapse in one hundred presumably healthy young females.

Clinical, electrocardiographic, phonocardiographic, and echocardiographic examinations were performed in 100 presumably healthy young females. Treadmill testing and ambulatory electrocardiographic monitoring were performed in a selected group of these subjects. Phonocardiograms, recorded with the subjects supine at rest, after inhalation of amyl nitrite, and in the upright position, revealed a ...

متن کامل

Evaluation of Björk-Shiley prosthetic valves by real-time two-dimensional Doppler echocardiographic flow mapping.

We studied the value of two-dimensional Doppler echocardiographic color flow mapping for identifying normal transvalve flow profiles and valve malfunction in 20 patients with Björk-Shiley prosthetic valves. Seven patients had Björk-Shiley prosthetic valves in the aortic position alone, seven in the mitral position, and six had prosthetic valves in both the aortic and mitral positions. In 10 pat...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Chest

دوره 70 2  شماره 

صفحات  -

تاریخ انتشار 1976