Comparison of optomagnetic and AC susceptibility readouts in a magnetic nanoparticle agglutination assay for detection of C-reactive protein.
نویسندگان
چکیده
There is an increasing need to develop biosensor methods that are highly sensitive and that can be combined with low-cost consumables. The use of magnetic nanoparticles (MNPs) is attractive because their detection is compatible with low-cost disposables and because application of a magnetic field can be used to accelerate assay kinetics. We present the first study and comparison of the performance of magnetic susceptibility measurements and a newly proposed optomagnetic method. For the comparison we use the C-reactive protein (CRP) induced agglutination of identical samples of 100nm MNPs conjugated with CRP antibodies. Both methods detect agglutination as a shift to lower frequencies in measurements of the dynamics in response to an applied oscillating magnetic field. The magnetic susceptibility method probes the magnetic response whereas the optomagnetic technique probes the modulation of laser light transmitted through the sample. The two techniques provided highly correlated results upon agglutination when they measure the decrease of the signal from the individual MNPs (turn-off detection strategy), whereas the techniques provided different results, strongly depending on the read-out frequency, when detecting the signal due to MNP agglomerates (turn-on detection strategy). These observations are considered to be caused by differences in the volume-dependence of the magnetic and optical signals from agglomerates. The highest signal from agglomerates was found in the optomagnetic signal at low frequencies.
منابع مشابه
Integration of Agglutination Assay for Protein Detection in Microfluidic Disc Using Blu-ray Optical Pickup Unit and Optical Fluid Scanning
We present a novel strategy for thrombin detection by combining a magnetic bead based agglutination assay and low-cost microfluidic disc. The detection method is based on an optomagnetic readout system implemented using a Blu-ray optical pickup unit (OPU) as main optoelectronic component. The assay, from sample to answer, is fully integrated on a microfluidic disc which embeds on-disc mixing en...
متن کاملMeasurement of C- Reactive Protein (CRP ) Using Magnetic Nanobeads Add ELISA
Abstract Background and Objective: Magnetic nanobeads have a large surface- area-to-volume ratio, which is used for immobilized antibody. Using nanoparticles could increase the amount of antibodies in surface in comparison to ELISA. We investigated the ability of magnetic nanobeads to evaluate CRP by colorimetric method and compared the results with ELISA. Material and Methods: This stud...
متن کاملOptomagnetic read-out enables easy, rapid, and cost-efficient qualitative biplex detection of bacterial DNA sequences
There is an increasing need to develop novel bioassay methods for low-cost, rapid, and easy-to-use multiplex detection of pathogens in various fields ranging from human infectious disease diagnosis, drinking water quality control, to food safety applications. Due to their unique advantages, magnetic and optomagnetic bioassay principles are particularly promising for biodetection platforms that ...
متن کاملTurn-on optomagnetic bacterial DNA sequence detection using volume-amplified magnetic nanobeads.
Detection of a Vibrio cholerae DNA-sequence using an optomagnetic read-out exploiting the dynamic behavior of magnetic nanobeads along with two turn-on data analysis approaches is demonstrated. The optomagnetic method uses a weak uniaxial AC magnetic field of varying frequency applied perpendicular to the optical path and measures the modulation of laser light passing through a cuvette containi...
متن کاملCharacterization of fine particles using optomagnetic measurements.
The remanent magnetic moment and the hydrodynamic size are important parameters for the synthesis and applications of magnetic nanoparticles (MNPs). We present the theoretical basis for the determination of the remanent magnetic moment and the hydrodynamic size of MNPs with a narrow size distribution using optomagnetic measurements. In these, the 2nd harmonic variation of the intensity of light...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Biosensors & bioelectronics
دوره 88 شماره
صفحات -
تاریخ انتشار 2017