Extension of the NIST AC-DC Difference Calibration Service for Current to 100 kHz

نویسندگان

  • Joseph R. Kinard
  • Thomas E. Lipe
  • Clifton B. Childers
چکیده

The NIST calibration service for ac-dc difference of thermal current converters relies on multijunction thermal converters as the primary standards, and various thermal converters and thermoelements (TEs) as the reference and working standards. Calibrations are performed by comparing the ac-dc difference of a customer's thermal current converter to the ac-dc difference of a NIST standard current converter. Typical artifacts accepted for calibration include single-junction thermoelements, multijunction thermal converters, and transfer shunts for use with TEs. This paper describes the standards on which the calibration service is based and the results of the study to characterize the NIST standards over the extended frequency range from 50 kHz to 100 kHz at currents from 1 mA to 20 A. The general method for the frequency extension at high frequency involves the use of thermoelements in the 5 mA range, with small frequency dependence, as the starting point for build-up and build-down chains to cover the whole range from 1 mA to 20 A.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Improvements in the Nist Calibration Service for Thermal Transfer Standards

We report on multiple improvements made in the calibration services offered for thermal converters and thermal transfer standards in the ac-dc Difference Standards and Measurement Techniques Project at NIST during the past year. The major improvement in this calibration is the increased efficiency made possible by the consolidation of three disparate calibration services – low voltage thermal t...

متن کامل

An AC Josephson Voltage Standard for AC-DC Transfer-Standard Measurements

In order to improve the National Institute of Standards and Technology (NIST) low-voltage calibration service for thermal-voltage converters, we have constructed an ac-Josephsonvoltage-standard (ACJVS) system that can synthesize accurate rms voltages of up to 100 mV. Using this system, we synthesized dc voltages and audio-frequency sine waves in order to measure ac–dc differences on the 22 mV a...

متن کامل

The Calibration of dc Voltage Standards at NIST

This document describes the procedures used at NIST to calibrate dc voltage standards in terms of the NIST volt. Three calibration services are offered by the Electricity Division: Regular Calibration Service (RCS) of client standard cells at NIST; the Volt Transfer Program (VTP) a process to determine the difference between the NIST volt and the volt as maintained by a group of standard cells ...

متن کامل

AC-DC Transfer Standard Measurements Using An AC Josephson Voltage Standard

NIST is implementing the Josephson arbitrary waveform synthesizer as an ac Josephson Voltage Standard (ACJVS). Over the past few years we have demonstrated precise measurements of synthesized sine waves at rms voltage amplitudes up to 170 mV. The system can generate a variety of precision voltage outputs including dc voltages and ac sinewaves, so it can be used as a quantum-based voltage source...

متن کامل

AC-DC Transfer Standard Measurements and Generalized Compensation with the AC Josephson Voltage Standard

We present ac-dc transfer standard measurements using the NIST pulse-driven ac Josephson voltage standard source. We have investigated the frequency dependence for several output voltages up to 200 mV for frequencies from 2.5 kHz to 100 kHz. We found that as the frequency increases, the ac-dc differences for the two arrays on the same chip do not agree. We investigated different attempts of imp...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 102  شماره 

صفحات  -

تاریخ انتشار 1997