Weight factors for limited angle photoacoustic tomography.

نویسندگان

  • G Paltauf
  • R Nuster
  • P Burgholzer
چکیده

Photoacoustic tomography (PAT) is based on the generation of ultrasound waves by heating an object with short light pulses. A three-dimensional image of the distribution of absorbed energy within the object is reconstructed from signals measured around the object with either point-like or extended, linear sensors. Limited angle artefacts arise when the curve or surface connecting neighbouring detectors is not closed around the object. For this case, there exists a 'detection region' in which all boundaries of an object are visible in the reconstruction. All straight lines passing through each point in this region intersect the detection curve or surface at least once. Although for these points an accurate reconstruction is possible, direct back projection leads to artefacts when some of the straight lines intersect the detection surface twice and others just once. In this work, special weight functions for direct, non-iterative back projection are presented that reduce these kinds of artefacts. A clear improvement in image quality is shown in simulations for three-dimensional (3D) imaging with point detectors and for two-dimensional (2D) imaging using line detectors compared to reconstruction without weight factors. For the 2D case also an experiment is shown. The presented weight factors make commonly used back projection formulae suitable for a more accurate reconstruction of the initial pressure distribution in cases where the detection aperture only covers a limited angle, and the region of interest lies within the detection region.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Photoacoustic image reconstruction from few-detector and limited-angle data

Photoacoustic tomography (PAT) is an emerging non-invasive imaging technique with great potential for a wide range of biomedical imaging applications. However, the conventional PAT reconstruction algorithms often provide distorted images with strong artifacts in cases when the signals are collected from few measurements or over an aperture that does not enclose the object. In this work, we pres...

متن کامل

Improving limited-view photoacoustic tomography with an acoustic reflector.

The versatility and real-time imaging capability of commercial linear array transducers make them widely used in clinical ultrasound and photoacoustic imaging. However, they often suffer from limited detection view. For instance, acoustic waves traveling at a grazing angle to the transducer surface are difficult to detect. In this letter, we propose a simple and easy approach to ameliorate this...

متن کامل

B-Mode Photoacoustic Imaging using Linear Array: Numerical Study for Forward-Backward Minimum Variance Beamformer Combined with Delay-Multiply-and-Sum

Photoacoustic imaging (PAI) is a promising medical imaging modality which provides the resolution of Ultrasound (US) and the contrast of Optical imaging modalities. One of the most important challenges in PAI is image formation, especially in the case that a linear-array US transducer is used for data acquisition. This is due to the fact that in the linear-array scenario, there is only 60 degre...

متن کامل

Tripling the detection view of high-frequency linear-array-based photoacoustic computed tomography by using two planar acoustic reflectors.

BACKGROUND Linear-array-based photoacoustic computed tomography (PACT) suffers from a limited view. Circular scanning does increase the detection view angle but is time-consuming. Therefore, it is desirable to increase the detection view angle of linear-array-based PACT without sacrificing imaging speed. METHODS Two planar acoustic reflectors placed at 120 degrees to each other were added to ...

متن کامل

A framework for directional and higher-order reconstruction in photoacoustic tomography.

Photoacoustic tomography is a hybrid imaging technique that combines high optical tissue contrast with high ultrasound resolution. Direct reconstruction methods such as filtered back-projection, time reversal and least squares suffer from curved line artefacts and blurring, especially in the case of limited angles or strong noise. In recent years, there has been great interest in regularised it...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Physics in medicine and biology

دوره 54 11  شماره 

صفحات  -

تاریخ انتشار 2009