Counting efficiency of some commercially available liquid scintillators compared.

نویسندگان

  • P Gray
  • O van Reenen
  • G M Potgieter
چکیده

We investigated four commercially available liquid scintillators as to overall counting efficiency and quenching effect of added phosphate-buffered isotonic saline. They differed significantly in these respects. Evidently, characteristics of the scintillator used in radioimmunoassays where beta-particle emitters are counted should be carefully examined if results are to be optimal.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Effect of Quench on Quantitating Alpha Radionuclides by Liquid Scintillation Counting

Detecting and quantitating alpha emitting radionuclides are routine tasks in nuclear energy and environmental monitoring. For this purpose, liquid scintillation counting (LSC) is an excellent method because alphas are counted with nearly 100% counting efficiency with low backgrounds, even with severe quenching. Quenching agents such as inorganic acids, organic acids, and extractive scintillator...

متن کامل

The Effect of Quench on Quantitating Alpha Radionuclides by Liquid Scintillation Counting

Detecting and quantitating alpha emitting radionuclides are routine tasks in nuclear energy and environmental monitoring. For this purpose, liquid scintillation counting (LSC) is an excellent method because alphas are counted with nearly 100% counting efficiency with low backgrounds, even with severe quenching. Quenching agents such as inorganic acids, organic acids, and extractive scintillator...

متن کامل

Comparison of the efficacy of biodegradable and non-biodegradable scintillation liquids on the counting of tritium- and [14C]-labeled compounds.

The widespread use of H and 14C in research has generated a large volume of waste mixed with scintillation liquid, requiring an effective control and appropriate storage of liquid radioactive waste. In the present study, we compared the efficacy of three commercially available scintillation liquids, Optiphase HiSafe 3, Ultima-Gold AB (biodegradable) and Insta-Gel-XF (non-biodegradable), in term...

متن کامل

Cocktails for Liquid Scintillation Counting

It is most important to assure 4 π−geometry in liquid scintillation counting for accurate results1. Usually this can be assured by dissolution of the sample in an appropriate cocktail. The cocktail does not only contain a solvent but also at least one scintillator and surfactants for aqueous samples because aqueous samples are completely insoluble in the pure organic solvents and scintillators....

متن کامل

Photostability Determination of Commercially Available Nifedipine Oral Dosage Forms in Iran

Nifedipine (NIF) a 1, 4-dihydropyridine calcium channel antagonist, undergoes photodegradation to nitroso analogues of dehydronifedipine (NDNIF) when exposed to sunlight. Photodegradation products of NIF have no clinical activity, so different formulations of NIF must remain unchanged. If NIF preparations become unstable in exposure to light, they could cause therapeutic failure. The present st...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Clinical chemistry

دوره 26 13  شماره 

صفحات  -

تاریخ انتشار 1980