Guidelines on preparation, certification, and use of certified plasmas for ISI calibration and INR determination--a rebuttal.
نویسندگان
چکیده
Reliable international normalized ratio (INR) determination depends on accurate values for international sensitivity index (ISI) and mean normal prothrombin time (MNPT). Local ISI calibration can be performed to obtain reliable INR. Alternatively, the laboratory may determine INR directly from a line relating local log(prothrombin time [PT]) to log(INR). This can be done by means of lyophilized or frozen plasmas to which certified values of PT or INR have been assigned. Currently there is one procedure for local calibration with certified plasmas which is a modification of the WHO method of ISI determination. In the other procedure, named 'direct' INR determination, certified plasmas are used to calculate a line relating log(PT) to log(INR). The number of certified plasmas for each procedure depends on the method of preparation and type of plasma. Lyophilization of plasma may induce variable effects on the INR, the magnitude of which depends on the type of thromboplastin used. Consequently, the manufacturer or supplier of certified plasmas must assign the values for different (reference) thromboplastins and validate the procedure for reliable ISI calibration or 'direct' INR determination. Certification of plasmas should be performed by at least three laboratories. Multiple values should be assigned if the differences between thromboplastin systems are greater than 10%. Testing of certified plasmas for ISI calibration may be performed in quadruplicate in the same working session. It is recommended to repeat the measurements on three sessions or days to control day-to-day variation. Testing of certified plasmas for 'direct' INR determination should be performed in at least three sessions or days. Correlation lines for ISI calibration and for 'direct' INR determination should be calculated by means of orthogonal regression. Quality assessment of the INR with certified plasmas should be performed regularly and should be repeated whenever there is a change in reagent batch or in instrument. Discrepant results obtained by users of certified plasmas should be reported to manufacturers or suppliers.
منابع مشابه
Simplified method for international normalized ratio (INR) derivation based on the prothrombin time/INR line: an international study.
BACKGROUND The need to perform local International Sensitivity Index (ISI) calibrations and in particular the requirement for a manual method for prothrombin time (PT) determination, have proved to be obstacles to application of the WHO scheme for PT standardization. METHODS We used international normalized ratio (INR) derived with a set of only 5 European Concerted Action on Anticoagulation ...
متن کاملNoninvasive prenatal diagnosis of a case of Down syndrome due to robertsonian translocation by massively parallel sequencing of maternal plasma DNA.
1. Remijn JA, Lucas S, Wildeboer B, van Suijlen JDE, Adriaansen HJ. Strongly increased international normalized ratio with recombinant Neoplastin R compared with tissue extract Neoplastin Plus in patients initiating oral anticoagulation therapy: implications for anticoagulation dosage. Clin Chem 2008;54:1929–31. 2. Roussi J, Drouet L, Samama M, Sié P. French multicentric evaluation of recombina...
متن کاملScreening INR deviation of local prothrombin time systems.
AIM To assess the reliability of local international normalised ratios (INR) using a set of three international reference preparation (IRP) certified freeze dried plasmas. METHODS 55 centres in the United Kingdom and the Republic of Ireland participated; 36 centres employed coagulometers and 19 a manual prothrombin time technique, all with the same batch of routine commercial thromboplastin. ...
متن کاملEvaluation of European Concerted Action on Anticoagulation lyophilized plasmas for INR derivation using the PT/INR line.
The prothrombin time/international normalized ratio (PT/INR) Line method based on 5 certified European Concerted Action on Anticoagulation (ECAA) plasmas provides reliable local INR values without conventional World Health Organization international sensitivity index calibrations. The present study investigated the use of different numbers and types of ECAA calibrant plasmas to derive accurate ...
متن کاملMulticentre ISI assignment and calibration of the INR measuring range of a new point-of-care system designed for home monitoring of oral anticoagulation therapy.
The new CoaguChek XS system is designed for use in patient selftesting. It is the successor of the current CoaguChek S system. The detection principle is based on the amperometric measurement of the thrombin activity initiated by starting the coagulation cascade using a human recombinant thromboplastin. This study was performed to assign the International SEnsitivity Index (ISI) to the new test...