OBJECTIVE To calibrate 1.5 tesla magnetic resonance scanners for the quantification of the concentration of iron in the liver. MATERIAL AND METHODS We analyzed twenty-eight 1.5 tesla magnetic resonance scanners using a phantom with four tubes containing different concentrations of iron (III) chloride and one tube without iron. The phantom represented two typical patients: one with moderate iron overload and one with high iron overload. We measured the signal intensity ratio between each iron-containing tube and the tube without iron; then we calculated the theoretical levels of iron concentration in each scanner according to the model for the two levels of overload. We compared the results of each scanner with those of the reference scanner in which the model and the phantom had been designed, and we calculated the percentage of difference between the two scanners. RESULTS The mean difference in the ratios compared to the reference center was 11% (0.3-39). The mean concentration of iron was 71 μmol Fe/g for moderate overload and 193 μmol Fe/g for high overload. The mean difference was 6% (1.2- 7%) and 3.4% (0-16%). respectively. In two scanners, we applied a correction factor so that the difference was below 25% in all cases. CONCLUSION We calibrated twenty-eight 1.5 tesla scanners for the concentration of iron in the liver and achieved variability less than 25%.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of Organ Specific Iron Overload in Transfusion-dependent Thalassemia by Magnetic Resonance Imaging Techniques

The consequence of repeated blood transfusions in thalassemia is iron overload in different organs. Magnetic resonance imaging (MRI) is a reliable, non-invasive and accurate method for iron detection in various tissues, hence the introduction of MRI has revolutionized the management of these patients and improved the life expectancy of them. Cardiac MRI T2* has a profound effect not only on est...

متن کامل

Cardiac and Hepatic T2*-Weighted Magnetic Resonance Imaging

Abstract Background Iron overload is the main transfusion related side effects in patients with transfusion dependent hemoglobinopathies. Severe iron deposition in tissues leads to organ dysfunction. Many organs can be affected such as heart, liver, and endocrine organs. Cardiac failure and liver fibrosis are the consequent of Iron overload in transfusion dependent hemoglobinopathy. Magne...

متن کامل

ANALYTICAL STUDY OF EFFECT OF BILAYER INORGANIC AND ORGANIC COATING AROUND THE IRON OXIDE NANOPARTICLES ON MAGNETIC RESONANCE IMAGING CONTRAST

Background & Aims: In recent years, iron oxide nanoparticles have been used in contrast-enhanced magnetic resonance imaging for diagnosing a wide range of diseases. In order to provide biocompatibility and prevent the toxicity of the nanoparticles, using organic or inorganic coating around these nanoparticles is important for their application. The aim of this study is to investigate the effect...

متن کامل

A New Potential Contrast Agent for Magnetic Resonance Imaging: Iron Oxide-4A Nanocomposite

Background: Magnetic resonance imaging (MRI) contrast agents have an important role to differentiate healthy and diseased tissues. Access and design new contrast agents for the optimal use of MRI are necessary. This study aims to evaluate iron oxide–4A nanocomposite ability to act as a magnetic resonance imaging contrast agent.Materials and Methods: Iron oxide–4A nanocomposite (F4A) was syn...

متن کامل

Static Coil Design Considerations for the Magnetic Resonance Imaging

One of the main challenges in developing magnetic resonance imaging (MRI) systems is to create a static coil that needs to generate magnetic field density along with the characteristics of optimal homogeneity and magnitude size. To do this, two N42 Block PMs are used and the iron core is designed and optimized in accordance with the dimensions of PM pieces using ANSYS Maxwell software. Then, al...

متن کامل

Comparison of Two Quantitative Susceptibility Mapping Measurement Methods Used For Anatomical Localization of the Iron-Incorporated Deep Brain Nuclei

Introduction Quantitative susceptibility mapping (QSM) is a new contrast mechanism in magnetic resonance imaging (MRI). The images produced by the QSM enable researchers and clinicians to easily localize specific structures of the brain, such as deep brain nuclei. These nuclei are targets in many clinical applications and therefore their easy localization is a must. In this study, we aimed to i...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Radiologia

دوره 54 2  شماره 

صفحات  -

تاریخ انتشار 2012