Accuracy and precision assessment for activity quantification in individualized dosimetry of 177Lu-DOTATATE therapy
نویسندگان
چکیده
BACKGROUND In order to obtain a reliable 177Lu-DOTATATE therapy dosimetry, it is crucial to acquire accurate and precise activity measurements with the radionuclide calibrator, the SPECT/CT camera, and the NaI(Tl) well counter. The aim of this study was to determine, in a clinical context, the accuracy and the precision of their activity quantification over a range of activities and time. Ninety-three 177Lu sources from the manufacturer were measured in the radionuclide calibrator over 2.5 years to evaluate its calibration accuracy and precision compared to the manufacturer's value. A NEMA 2012/IEC 2008 phantom was filled with a 177Lu activity concentration sphere-to-background ratio of five. It was acquired with the SPECT/CT camera to determine the reconstruction parameters offering the best compromise between partial volume effect and signal-to-noise ratio. The calibration factor was computed accordingly. The calibration quality was monitored over 2.5 years with 33 phantom acquisitions with activities ranging from 7040 to 0.6 MBq. Home-made sources were used to calibrate the well counter. Its reliability was evaluated with activities ranging from 150 to 0.2 kBq measured 34 times over 2.5 years. RESULTS For the radionuclide calibrator, median [interquartile range] for the error on activity measurement was -0.99 [1.31] %. The optimal SPECT reconstruction parameters were obtained with 16 iterations, 16 subsets and a 12-mm Gaussian post-filter. The calibration factor was 9.87 cps/MBq with an error of -1.05 [2.12] %. The well counter was calibrated with 31.5 cps/kBq, and the error was evaluated to -12.89 [16.55] %. CONCLUSIONS The accuracy and the precision of activity quantification using dedicated quality control were found to be sufficient for use in dosimetry implemented in clinical routine. The proposed methodology could be implemented in other centres to obtain reproducible 177Lu-based treatment dosimetry.
منابع مشابه
A comparison of 2D and 3D kidney absorbed dose measures in patients receiving 177Lu-DOTATATE
Objective(s): To investigate and compare quantitative accuracy of kidney absorbed dose measures made from both 2D and 3D imaging in patients receiving 177LuDOTATATE (Lutate) for treatment of neuroendocrine tumours (NETs). Methods: Patients receiving Lutate therapy underwent both whole body planar imaging and SPECT/CT imaging over the kidneys at time points 0.5, 4, 24, and 96-120 hours after inj...
متن کاملThe comparison of serial SPECT-CT imaging to estimate absorbed dose to the organ at risk from peptide receptor radionuclide therapy dosimetry
Introduction: In Peptide Receptor Radionuclide Therapy (PRRT), the administration of radionuclide such as Lu-177 label with a pharmaceutical agent useful to destroy the lesion. The amount of Lu-177 radioactivity administered to the patients is still not standardize and generally not more than 7.4 GBq per session due to the patient’s safety issues. The first cycle of Lu-177 is a...
متن کاملSPECT image segmentation for estimation of tumour volume and activity concentration in 177Lu-DOTATATE radionuclide therapy
BACKGROUND Dosimetry in radionuclide therapy has the potential to allow for a treatment tailored to the individual patient. One therapeutic radiopharmaceutical where patient-specific dosimetry is feasible is 177Lu-DOTATATE, used for the treatment of neuroendocrine tumours. The emission of gamma photons by 177Lu allows for imaging with SPECT (single photon emission computed tomography). One impo...
متن کاملDemonstration of focal physiologic in-vivo somatostatin receptor expression in the caput epididymis of the testes on 68Ga-DOTANOC PET/CT and 177Lu-DOTATATE post-therapy whole body scintigraphy
We present the case of a 60-year-old man with metastatic neuroendocrine tumor of the ileum following ileal resection, being evaluated for 177Lu-based peptide receptor radionuclide therapy. 68Ga-DOTANOC PET/CT showed focal increased tracer uptake in the scrotal region without any morphologic changes on the corresponding CT images. Similar increased tracer uptake was seen on...
متن کاملLutetium-177 DOTATATE Production with an Automated Radiopharmaceutical Synthesis System
Objective(s): Peptide Receptor Radionuclide Therapy (PRRT) with yttrium-90 (90Y) and lutetium-177 (177Lu)-labelled SST analogues are now therapy option for patients who have failed to respond to conventional medical therapy. In-house production with automated PRRT synthesis systems have clear advantages over manual methods resulting in increasing use in hospital-based radiopharmacies. We report...