Simultaneous Reduction in Noise and Cross-Contamination Artifacts for Dual-Energy X-Ray CT

نویسندگان

  • Baojun Li
  • Baohong Li
  • Jack Luo
  • Peng Tang
  • Jiandong Mao
  • Xiaoye Wu
چکیده

PURPOSE Dual-energy CT imaging tends to suffer from much lower signal-to-noise ratio than single-energy CT. In this paper, we propose an improved anticorrelated noise reduction (ACNR) method without causing cross-contamination artifacts. METHODS The proposed algorithm diffuses both basis material density images (e.g., water and iodine) at the same time using a novel correlated diffusion algorithm. The algorithm has been compared to the original ACNR algorithm in a contrast-enhanced, IRB-approved patient study. Material density accuracy and noise reduction are quantitatively evaluated by the percent density error and the percent noise reduction. RESULTS Both algorithms have significantly reduced the noises of basis material density images in all cases. The average percent noise reduction is 69.3% and 66.5% with the ACNR algorithm and the proposed algorithm, respectively. However, the ACNR algorithm alters the original material density by an average of 13% (or 2.18 mg/cc) with a maximum of 58.7% (or 8.97 mg/cc) in this study. This is evident in the water density images as massive cross-contaminations are seen in all five clinical cases. On the contrary, the proposed algorithm only changes the mean density by 2.4% (or 0.69 mg/cc) with a maximum of 7.6% (or 1.31 mg/cc). The cross-contamination artifacts are significantly minimized or absent with the proposed algorithm. CONCLUSION The proposed algorithm can significantly reduce image noise present in basis material density images from dual-energy CT imaging, with minimized cross-contaminations compared to the ACNR algorithm.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Computed tomography based attenuation correction in PET/CT: Principles, instrumentation, protocols, artifacts and future trends

  The advent of dual-modality PET/CT imaging has revolutionized the practice of clinical oncology, cardiology and neurology by improving lesions localization and the possibility of accurate quantitative analysis. In addition, the use of CT images for CT-based attenuation correction (CTAC) allows to decrease the overall scanning time and to create a noise-free attenuat...

متن کامل

Compensation of Cross-Contamination in Simultaneous 201Tl/99mTc Myocardial Perfusion SPECT Imaging

Introduction: It is a common protocol to use 201Tl for the rest and 99mTc for the stress cardiac SPECT imaging. Theoretically, both types of imaging may be performed simultaneously using different energy windows for each radionuclide. However, a potential limitation is the cross-contamination of scattered photons from 99mTc and collimator X-rays into the 201Tl energy window. We used a middle en...

متن کامل

Shearlet-Based Adaptive Noise Reduction in CT Images

The noise in reconstructed slices of X-ray Computed Tomography (CT) is of unknown distribution, non-stationary, oriented and difficult to distinguish from main structural information. This requires the development of special post-processing methods based on the local statistical evaluation of the noise component. This paper presents an adaptive method of reducing noise in CT images employing th...

متن کامل

A Model-Based Iterative Algorithm for Dual-Energy X-Ray CT Reconstruction

Recent developments in dual-energy X-ray CT have shown a number of benefits over standard CT for object separation, contrast enhancement, artifact reduction, and material composition assessment. As with traditional CT, model-based iterative approaches to reconstruction offer the opportunity to reduce noise and artifacts in dual energy reconstructions. However, previous approaches to model-based...

متن کامل

Comparing X-ray dose reduction capability of silicon-bismuth micro- and nanocomposite shields using chest CT test

Comparison of radiation shields of different thicknesses and structures (especially different particle sizes of their constituents) is helpful to select an appropriate shield for patient protection in medical imaging tests particularly CT-scan. In this study micro- and nano-powders of bismuth were mixed with silicone matrix to make composite radiation shields for breast. The shields were provid...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 2013  شماره 

صفحات  -

تاریخ انتشار 2013