Fatigue attributable to broken pacemaker header by baseball.
نویسندگان
چکیده
A 58-year-old male baseball coach with a prepectoral dualchamber permanent pacemaker (Accent DR 2210; St. Jude Medical) implanted for sinus node dysfunction was evaluated for a pocket hematoma after he was hit by a baseball over the pacemaker. Data from the pacemaker showed normal atrial and ventricular capture, but both lead impedances were elevated (although still within the manufacturer’s normal limits). Brief episodes of noise were seen on the ventricular channel. The hematoma resolved, but 4 months later, the patient presented with dizziness and fatigue of 1 week’s duration. A chest radiograph did not reveal any lead or pacemaker generator abnormalities. Pacemaker data showed sequential atrial and ventricular pacing at 60 beats per minute (Figure A; marker, atrial, and ventricular channels). However, the right ventricular lead capture threshold was abnormal, and impedance had increased further. Simultaneous surface electrocardiography showed sinus bradycardia at 37 beats per minute (Figure A; channel 1). Opening the pocket revealed complete detachment of the pacemaker header from the rest of the generator (Figure B and C). The patient had significant subcutaneous adipose tissue (body mass index, 35.8 kg/m). Symptoms subsided after the pacemaker generator was replaced.
منابع مشابه
The Evaluation of Wheat Losses Harvesting in Two Conventional Combine (John Deere 1165, 955) in Iran
Wheat is one of the most important food staff in consumption pattern of each country. More than 50 percent of energy is provided by bread in the developing countries. The aim of this study was measuring wheat losses during pre-harvest and harvest stages in Chaharmahal and Bakhtiari province of Iran in year 2013. Wheat losses at harvest stages were measured to wheat varieties in 2 levels of Omid...
متن کاملThe Evaluation of Wheat Losses Harvesting in Two Conventional Combine (John Deere 1165, 955) in Iran
Wheat is one of the most important food staff in consumption pattern of each country. More than 50 percent of energy is provided by bread in the developing countries. The aim of this study was measuring wheat losses during pre-harvest and harvest stages in Chaharmahal and Bakhtiari province of Iran in year 2013. Wheat losses at harvest stages were measured to wheat varieties in 2 levels of Omid...
متن کاملPacemaker syndrome with sub-acute left ventricular systolic dysfunction in a patient with a dual-chamber pacemaker: consequence of lead switch at the header
In the daily practice of pacemaker insertion, the occurrence of atrial and ventricular lead switch at the pacemaker box header is a rare and unintentional phenomenon, with less than five cases reported in the literature. The lead switch may have dire consequences, depending on the indication for the pacemaker. One of these consequences is pacemaker syndrome, in which the normal sequence of atri...
متن کاملPacemaker Failure Due to Loss of Fluid Seal in a Patient with a 5/6mm Pacemaker Header Port and a 5mm Unipolar Lead
INTRODUCTION Before IS-1 (3.2 mm) standardization of pacemaker leads and connectors, 5/6 mm connector ports accomodated 5mm or 6mm diameter lead connector pins. CASE REPORT A patient with sick sinus syndrome underwent implantation of a 5mm unipolar atrial lead, mated to a 5/6mm connector port Medtronic Spectrax Sx 5985 pacemaker. Pulse generator reached ERI in 2006, with change out to a Medtr...
متن کاملEpidemiology of Shoulder and Elbow Injuries in Iranian Baseball Premier League
Background: Baseball is a team sport that has a total of nine main players and includes explosive, endurance, and strength movements that can cause multiple injuries over a season due to repetitive movements, long training, and competition. The purpose of the present study was to investigate the epidemiology of shoulder and elbow injuries in Iranian Premier League baseball players. Methods: 79...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Circulation. Arrhythmia and electrophysiology
دوره 6 5 شماره
صفحات -
تاریخ انتشار 2013