ToF performance evaluation of a PET insert with Digital Silicon Photomultiplier technology during MR operation

نویسندگان

  • David Schug
  • Jakob Wehner
  • Peter Michael Dueppenbecker
  • Bjoern Weissler
  • Pierre Gebhardt
  • Benjamin Goldschmidt
  • Torsten Solf
  • Volkmar Schulz
چکیده

Department of Physics of Molecular Imaging Systems, Institute for Experimental Molecular Imaging, RWTH Aachen University, Aachen, Germany 2012, our group presented the Hyperion IID platform which uses PDPC’s digitial SiPMs (DPC). In this work we use the same platform equipped with scintillator dimensions closer to a clinical application. This allows an investigation of the time of flight (ToF) performance of the platform and its behavior during simultaneous MR operation. We employ LYSO crystal arrays of 4×4×10 mm3 coupled to 4×4 PDPC DPC 3200-22 sensors allowing a one-to-one coupling of crystals to readout channels. Six sensor stacks are mounted onto a singles processing unit in a 2×3 arrangement. Two units are mounted on a gantry with a diameter of 216 mm. The DPCs are cooled down to approximately 5-10 °C under operation. We disable 20% of the worst cells and use an overvoltage (OV) of 2.0V and 2.5V. To obtain the best time stamps we use the first photon trigger and employ data quality cuts to filter out crystal scatter events. A narrow energy window of 511±50 keV is used and a minimal light fraction of the main pixel of more than 65% is requested. Using a Na22 point source in the isocenter of the modules the coincidence resolution time (CRT) of the two modules is evaluated outside the MR and inside the MR using different MR sequences. Gradient stress tests with switching z-gradients are performed. Inside the B0 field at 2.0V overvoltage the energy resolution is 11.45% (FWHM) and the CRT is 250ps (FWHM). At 2.5V overvoltage, the energy resolution is 11.15% (FWHM) and the CRT is 240ps (FWHM). During the heavy z-gradient sequence the energy resolution is degraded by 4.1% at 2V and 9.2% at 2.5V. The degradation of the CRT is 25% at 2V and 52% at 2.5V OV. During standard TSE and EPI sequences the CRT and energy resolution is not influenced. The Hyperion IID platform proofs to deliver good timing performance outside and inside an MR. The CRT is not influenced (<0.2%) using normal MR imaging sequences. Schug et al. EJNMMI Physics 2014, 1(Suppl 1):A20 http://www.ejnmmiphys.com/content/1/S1/A20

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of PET performance and MR compatibility of a preclinical PET/MR insert with digital silicon photomultiplier technology

Department of Physics of Molecular Imaging Systems, RWTH Aachen University, Germany In this work we present detailed characterizations of our preclinical high resolution PET/ MR insert based on the Hyperion-IID platform. The PET/MR insert consists of a ring of 10 singles detection modules, each comprising 2×3 scintillation detector stacks. Each detector stack features a 30×30 pixelated LYSO cry...

متن کامل

Performance of a high sensitivity time-of-flight PET ring operating simultaneously within a 3T MR system

Stanford University, USA A time-of-flight (TOF)-PET/MR research system installed at Stanford will be used to test the hypotheses that (a) it is possible to acquire simultaneous TOF-PET and 3T MR data while achieving uncompromised performance in both modalities and (b) simultaneous TOF-PET/MR is a tool for multi-parameter characterization of disease. In this paper we will describe the design as ...

متن کامل

PET performance and MRI compatibility evaluation of a digital, ToF-capable PET/MRI insert equipped with clinical scintillators.

We evaluate the MR compatibility of the Hyperion-II(D) positron emission tomography (PET) insert, which allows simultaneous operation in a clinical magnetic resonance imaging (MRI) scanner. In contrast to previous investigations, this work aims at the evaluation of a clinical crystal configuration. An imaging-capable demonstrator with an axial field-of-view of 32 mm and a crystal-to-crystal spa...

متن کامل

PET/MRI insert using digital SiPMs: Investigation of MR-compatibility

In this work, we present an initial MR-compatibility study performed with the world's first preclinical PET/MR insert based on fully digital silicon photo multipliers (dSiPM). The PET insert allows simultaneous data acquisition of both imaging modalities and thus enables the true potential of hybrid PET/MRI. Since the PET insert has the potential to interfere with all of the MRI's subsystems (s...

متن کامل

Recent development in PET instrumentation.

Positron emission tomography (PET) is used in the clinic and in vivo small animal research to study molecular processes associated with diseases such as cancer, heart disease, and neurological disorders, and to guide the discovery and development of new treatments. This paper reviews current challenges of advancing PET technology and some of newly developed PET detectors and systems. The paper ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 1  شماره 

صفحات  -

تاریخ انتشار 2014