Application of a test package in an intercomparison of the photon dose calculation performance of treatment planning systems used in a clinical setting.
نویسندگان
چکیده
BACKGROUND AND PURPOSE Testing the performance of treatment planning systems by using the AAPM Task Group 23 test package is a useful approach, but has its limitations. To be able to include technical developments, such as the asymmetric collimator, it was decided to remeasure the AAPM data set on more modern radiotherapy equipment, to extend the test geometries, and to evaluate the use of the new package. MATERIALS AND METHODS A coherent set of beam data of 6, 10 and 18 MV photon beams was measured on two modern linear accelerators. These data served as input data in seven commercially available treatment planning systems, which were clinically in use in different radiotherapy departments. Next, a test package was measured which included a missing tissue geometry and fields with asymmetrical collimator setting, with and without a wedge. RESULTS The absolute dose prediction from the different treatment planning systems in which the measured beam data were entered, was compared for all test points with the results of direct measurements. The criteria of acceptability were exceeded by some systems in cases of irregular field geometry and missing tissue geometry. The majority of the systems had difficulties with accurate dose calculation for asymmetrically wedged fields. CONCLUSIONS The application of the new test package did not introduce insuperable difficulties and was highly appreciated by the participating centres. Most systems performed reasonably well for the majority of the beam geometries, with the exception of asymmetrically wedged beams. The extended test package is available for other users or user groups for the purpose of commissioning new treatment planning systems, or new releases of existing systems.
منابع مشابه
Dose Calculations for Lung Inhomogeneity in High-Energy Photon Beams and Small Beamlets: A Comparison between XiO and TiGRT Treatment Planning Systems and MCNPX Monte Carlo Code
Introduction Radiotherapy with small fields is used widely in newly developed techniques. Additionally, dose calculation accuracy of treatment planning systems in small fields plays a crucial role in treatment outcome. In the present study, dose calculation accuracy of two commercial treatment planning systems was evaluated against Monte Carlo method. Materials and Methods Siemens Once or linea...
متن کاملEvaluation of Three-dimensional Treatment Planning System (TPS) performance in dose calculation of virtual wedged fields using film dosimetry
Introduction: Nowadays radiotherapy plays an important role in cancer treatment. Different radiotherapy techniques improvement emphasizes on using of the precise ، appropriate and useful algorithms. one of these techniques are wedged which is used in radiotherapy to compensate missing tissues and create a uniform dose distribution in tissues. The Siemens Artiste linear accelera...
متن کاملDosimetric Evaluation of Dose calculation algorithms of Monaco Treatment Planning System in the heterogeneities area
Introduction: In radiation therapy, the accuracy of dose calculations by a treatment planning system (TPS) is important to achieve tumor control and to spare normal tissue. Treatment planning system calculations in the heterogeneous situation may present significant inaccuracies. In this study, three different dose calculation algorithms, pencil beam (PB), collapsed cone (CC), ...
متن کاملValidation of treatment planning system using simulation PRIMO code.
Introduction: In radiation therapy, in order to double-check the dosimetric results of the main treatment planning system (TPS), a distinct TPS, with few capacitances in terms of contouring and a variety of dose calculation algorithms is used. This system has the capability to double check the planification and the accurate prediction of dose distribution in order to be ensured...
متن کاملEvaluation of CorePlanTM treatment planning system for 6 and 15 MV photon irradiation
Introduction: Radiotherapy is an effective modality to treat esophageal carcinoma. Due to various inhomogenous tissues with a variety of physical and radiological properties which can cause uncertainty in accurate dose calculations in radiation therapy, accuracy of TPS is very important. The aim of this study was to quantify dose calculation accuracy of CoreplanTM TPS in an inh...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Radiotherapy and oncology : journal of the European Society for Therapeutic Radiology and Oncology
دوره 60 2 شماره
صفحات -
تاریخ انتشار 2001