The Calibration of BEPC Beam Position Monitors

نویسندگان

  • K. Ye
  • H. Huang
چکیده

The basic requirement for the BEPC beam position monitor is the measurement of the beam orbit with 0.1 mm precision near the collision point. To improve the measurement accuracy, the response of the beam position monitor pickups was mapped in the laboratory before they were installed in the BEPC ring. The microcomputer-controlled test set consists of high frequency coaxial switches to select each pickup electrode, a movable antenna to simulate the beam, a signal source, a spectrum analyzer to measure the pickup signals, and analysis software. The signal source operates well below the -3 dB cutoff frequency of the pickups (buttons). We believe that the low-frequency measurement yields the same information as the real beam. The button signals were clear. This calibration technique is satisfactory for BEPC operation. INTRODUCTION Four-button type beam position monitors (BPM) are used in the BEPC 2.2 GeV storage ring. The BPM assembly is an electrostatic type with four disk electrodes and BNC vacuum feedthrough connectors. The beam pipe is a cylinder. The buttons are rotated 45 degrees off vertical and horizontal axes to avoid the fan of synchrotron radiation. A tooling ball located on the top of the vacuum chamber is used as the fiducial mark for survey and alignment of the BPM assembly, relative to an adjacent quadrupole magnet. Electrical differences in the buttons and the mechanical installation tolerances cause the BPM to report beam offsets that are not real. We measured these offsets in our test set. The button disk is made of stainless steel. The disks are welded to the center conductor of the BNC feedthroughs. The feedthroughs themselves are welded into the vacuum pipe. The disks are flush with the vacuum pipe wall. Button signals can reach a few GHz. A spectrum analyzer was used to measure button response. The BPM quality was measured with three different tests. First, we measured the button response to a fast pulse on an antenna in the center of the chamber. This measurement gave the button sensitivity to beam current. In the second test, we excited the antenna with a low-frequency field with the antenna centered. This gave us the BPM offset due to electrical and mechanical errors. If the buttons had equal capacitance and were perfectly installed relative to the center of the beam pipe, there would be no offset error and all buttons would produce identical signals. The third test we performed was an evaluation of the BPM sensitivity to antenna position. Figure 1 shows button signals from a typical antenna scan. FIGURE 1. Button outputs over 10 X 10 mm scan. The measuring setup is shown schematically in Figure 2. The antenna may be moved transversely inside the monitor along the Xand Y-axes. A SP6T rf switch is used to select signals from each button. The insertion loss of the switch was tested and corrections were made to the data. The attenuation of all button cables were measured. The mechanical error of the stepping motor was considered while the antenna moved. A personal computer (PC) controls the equipment via a PC I/O board, the RS232 port, and a GPIB board. The measurement is completely automatic. wGenerator BPM Coaxial E-104 Switch AmpPfk FIGURE 2. BPM calibration method schematic. The button electrode faces a beam, and the button senses an image current lb: * 2!c2&% b dt b dt (1) where p is the linear charge density, a is the radius of the button, and b is the radius of the duct (beam pipe). The self-capacitance of the button to the wall of the beam duct is Cb. A load resistor R, in shunt with C,, will produce a frequency-dependent coupling impedance to the beam that acts like a high-pass filter: The button output signal is as follows: v=Ibz=I R b 1 + iwRCb R l+iwRCb )I@) =$ ; )Rl(w) (2)

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The beam energy calibration system for the BEPC - II collider

This document contains a proposal of the BEPC-II collider beam energy calibration system (IHEP, Beijing). The system is based on Compton backscattering of carbon dioxide laser radiation, producing a beam of high energy photons. Their energy spectrum is then accurately measured by HPGe detector. The high-energy spectrum edge will allow to determine the average electron or positron beam energy ε ...

متن کامل

Beam-based Offset Calibration of the Pls Bpm

There are 108 beam position monitors in the PLS for the orbit measurement and closed orbit correction. As the BPM pickup electrodes are assembled directly on the 10m-long vacuum chamber, calibration of the electrical center of BPMs have been done with the external rf transmission-reflection method. Since the closed orbit distortion is still larger than 500 mm rms, we are developing the beam-bas...

متن کامل

Synchrotron Radiation Monitoring System at Bepc

The synchrotron radiation by the relativistic particles is a useful and important diagnostic tool to obtain various beam parameters. The visible light of the synchrotron radiation emitted in a bending magnet of the BEPC storage ring is reflected through the window by a watercooled mirror in the vacuum chamber and then directed by a remote controlled mirror onto the optical bench supporting the ...

متن کامل

Secondary Beam Monitors for the NuMI Facility at FNAL

The Neutrinos at the Main Injector (NuMI) facility is a conventional neutrino beam which produces muon neutrinos by focusing a beam of mesons into a long evacuated decay volume. We have built four arrays of ionization chambers to monitor the position and intensity of the hadron and muon beams associated with neutrino production at locations downstream of the decay volume. This article describes...

متن کامل

Recent Studies on Beam-photoelectron Instability in Bepc

An experiment investigating the photoelectron instability has been performed in BEPC at IHEP, China. The betatron oscillation of each bunch was observed and recorded by a single pass beam position monitor system which has a capability to measure the transverse position of every bunch over 16000 turns. The effect of an octupole magnet on the instability was also studied. This paper describes res...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 1998