Commissioning and initial experience with the first clinical gantry‐mounted proton therapy system
نویسندگان
چکیده
The purpose of this study is to describe the comprehensive commissioning process and initial clinical experience of the Mevion S250 proton therapy system, a gantry-mounted, single-room proton therapy platform clinically implemented in the S. Lee Kling Proton Therapy Center at Barnes-Jewish Hospital in St. Louis, MO, USA. The Mevion S250 system integrates a compact synchrocyclotron with a C-inner gantry, an image guidance system and a 6D robotic couch into a beam delivery platform. We present our commissioning process and initial clinical experience, including i) CT calibration; ii) beam data acquisition and machine characteristics; iii) dosimetric commissioning of the treatment planning system; iv) validation through the Imaging and Radiation Oncology Core credentialing process, including irradiations on the spine, prostate, brain, and lung phantoms; v) evaluation of localization accuracy of the image guidance system; and vi) initial clinical experience. Clinically, the system operates well and has provided an excellent platform for the treatment of diseases with protons.
منابع مشابه
Commissioning of the world's first compact pencil‐beam scanning proton therapy system
This paper summarizes clinical commissioning of the world's first commercial, clinically utilized installation of a compact, image-guided, pencil-beam scanning, intensity-modulated proton therapy system, the IBA Proteus® ONE, at the Willis-Knighton Cancer Center (WKCC) in Shreveport, LA. The Proteus® ONE is a single-room, compact-gantry system employing a cyclotron-generated proton beam with im...
متن کاملCommissioning and initial stereotactic ablative radiotherapy experience with Vero
The purpose of this study is to describe the comprehensive commissioning process and initial clinical performance of the Vero linear accelerator, a new radiotherapy device recently installed at UT Southwestern Medical Center specifically developed for delivery of image-guided stereotactic ablative radiotherapy (SABR). The Vero system utilizes a ring gantry to integrate a beam delivery platform ...
متن کاملMonte Carlo simulation of secondary neutron dose for scanning proton therapy using FLUKA
Proton therapy is a rapidly progressing field for cancer treatment. Globally, many proton therapy facilities are being commissioned or under construction. Secondary neutrons are an important issue during the commissioning process of a proton therapy facility. The purpose of this study is to model and validate scanning nozzles of proton therapy at Samsung Medical Center (SMC) by Monte Carlo simu...
متن کاملCommissioning and early experience with a new generation low-energy linear accelerator in zahedan
Introduction: Since commissioning beam data are treated as a reference and ultimately used by treatment planning systems, it is vitally important that the collected data are of the highest quality to avoid dosimetric and patient treatment errors that may subsequently lead to a poor radiation outcome. Beam data commissioning should be performed with appropriate knowledge and pr...
متن کاملBeam Diagnostics for the Proton Therapy Facility Proscan
The dedicated new medical facility PROSCAN has restarted the treatment of deep seated tumours with proton beams at Gantry 1. The beam lines to the new gantry 2 for the further development of the spot-scanning technology, as well as to the replacement of the long established OPTIS facility for the treatment of eye melanoma, and to a material irradiation area, are in the commissioning phase. An o...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 17 شماره
صفحات -