Assessing the Risks Associated with MRI in Patients with a Pacemaker or Defibrillator.

نویسندگان

  • Robert J Russo
  • Heather S Costa
  • Patricia D Silva
  • Jeffrey L Anderson
  • Aysha Arshad
  • Robert W W Biederman
  • Noel G Boyle
  • Jennifer V Frabizzio
  • Ulrika Birgersdotter-Green
  • Steven L Higgins
  • Rachel Lampert
  • Christian E Machado
  • Edward T Martin
  • Andrew L Rivard
  • Jason C Rubenstein
  • Raymond H M Schaerf
  • Jennifer D Schwartz
  • Dipan J Shah
  • Gery F Tomassoni
  • Gail T Tominaga
  • Allison E Tonkin
  • Seth Uretsky
  • Steven D Wolff
چکیده

BACKGROUND The presence of a cardiovascular implantable electronic device has long been a contraindication for the performance of magnetic resonance imaging (MRI). We established a prospective registry to determine the risks associated with MRI at a magnetic field strength of 1.5 tesla for patients who had a pacemaker or implantable cardioverter-defibrillator (ICD) that was "non-MRI-conditional" (i.e., not approved by the Food and Drug Administration for MRI scanning). METHODS Patients in the registry were referred for clinically indicated nonthoracic MRI at a field strength of 1.5 tesla. Devices were interrogated before and after MRI with the use of a standardized protocol and were appropriately reprogrammed before the scanning. The primary end points were death, generator or lead failure, induced arrhythmia, loss of capture, or electrical reset during the scanning. The secondary end points were changes in device settings. RESULTS MRI was performed in 1000 cases in which patients had a pacemaker and in 500 cases in which patients had an ICD. No deaths, lead failures, losses of capture, or ventricular arrhythmias occurred during MRI. One ICD generator could not be interrogated after MRI and required immediate replacement; the device had not been appropriately programmed per protocol before the MRI. We observed six cases of self-terminating atrial fibrillation or flutter and six cases of partial electrical reset. Changes in lead impedance, pacing threshold, battery voltage, and P-wave and R-wave amplitude exceeded prespecified thresholds in a small number of cases. Repeat MRI was not associated with an increase in adverse events. CONCLUSIONS In this study, device or lead failure did not occur in any patient with a non-MRI-conditional pacemaker or ICD who underwent clinically indicated nonthoracic MRI at 1.5 tesla, was appropriately screened, and had the device reprogrammed in accordance with the prespecified protocol. (Funded by St. Jude Medical and others; MagnaSafe ClinicalTrials.gov number, NCT00907361 .).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Pacemaker lead endocarditis with hiccups (Kalayci)

Background: Lead-related infections that might develop after pacemaker implantation associated with high mortality and morbidity rates are challenging to manage and pose high-cost. Patients with lead-related infections usually present with fever, chills and fatigue and the treatment can be challenging unless the implant system is extracted. Case presentation: A 66-year old male patient who und...

متن کامل

Food and Drug Administration perspective: Magnetic resonance imaging of pacemaker and implantable cardioverter-defibrillator patients.

In April 2005, the US Food and Drug Administration (FDA) commented on the current research regarding the use of magnetic resonance imaging (MRI) on patients with a pacemaker or implantable cardioverter-defibrillator (ICD).1 Pacemaker and ICD labeling currently cautions physicians against the use of MRI, and MRI manufacturers contraindicate MRI for pacemaker and ICD patients. As discussed in our...

متن کامل

Is correction for metallic artefacts mandatory in cardiac SPECT/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?

Introduction: Metallic artifacts due to pacemaker/ implantable cardioverter defibrillator (ICD) leads in CT images can produce artifactual uptake in cardiac SPECT/CT images. The aim of this study was to determine the influence of the metallic artifacts due to pacemaker and ICD leads on myocardial SPECT/CT imaging. Methods: The study included 9 patients wh...

متن کامل

Magnetic resonance imaging safety in pacemaker and implantable cardioverter defibrillator patients: how far have we come?

Magnetic resonance imaging (MRI) has long been regarded a general contraindication in patients with cardiovascular implanted electronic devices such as cardiac pacemakers or cardioverter defibrillators (ICDs) due to the risk of severe complications and even deaths caused by interactions of the magnetic resonance (MR) surrounding and the electric devices. Over the last decade, a better understan...