Implantation of a Modified Baerveldt Glaucoma Implant with a Smaller Tube and Intraluminal Stent
نویسندگان
چکیده
Dear Editor, The Baerveldt glaucoma implant (BGI; Abbott Laboratories, Abbott Park, IL, USA) is a widely used glaucoma drainage device for intraocular pressure (IOP) control. Although the BGI effectively controls IOP, the possibility of postoperative ocular hypotony owing to excessive aqueous drainage and tube-related complications still remains [1,2]. Given that the amount of aqueous drainage at the early postoperative period is mainly determined by tube profiles, we reasoned that a tube with a smaller diameter and intraluminal stent may have a lower risk of postoperative ocular hypotony [3-5]. In addition, a smaller tube may have a lower chance of tube-related complications, such as conjunctival erosion or tube exposure [3-5]. To test this, we replaced the conventional silicone tube of the BGI (630-μm external and 300-μm internal diameter) with a smaller silicone tube (300-μm external and 200-μm internal diameter) and an intraluminal stent (5-0 nylon); we named this tube “Finetube [4].” We implanted the modified BGI with a Finetube into three eyes from three patients with uncontrolled IOP (36, 34, and, 30 mmHg, respectively) despite previous trabeculectomies, as well as medical treatment. Preoperatively, the conventional BGI tube was cut approximately 1 mm from the body plate. The Finetube was inserted into the stump of the conventional tube and fixed using silicone adhesive (Fig. 1A). An intraluminal stent prevents excessive aqueous drainage through the tube, and induces additional IOP reduction after surgery by retracting the stent (Fig. 1B). The main conjunctival incision was made 8 mm posterior to the limbus and the BGI body plate was placed under the superior and lateral recti muscles and fixed to the bared sclera, 10 mm posterior to the limbus. The second conjunctival incision was made 2 mm posterior to the limbus. A scleral tunnel, from the limbal second incision to the forniceal main incision was made using a 23-gauge needle. The needle tip was used to guide the Finetube towards the second incision. After the tube was trimmed to an appropriate length, it was inserted into the anterior chamber through the corneoscleral track created using a 26-gauge needle as described previously [4]. The corneoscleral track and conjunctiva were closed using 10-0 nylon sutures. A part of the 5-0 nylon stent was left partially exposed through the conjunctiva to be pulled out when the IOP needed to be further reduced [4]. At the early postoperative period (within 1 month), the IOP ranged from 6 to 15 mmHg; no eye showed postoperative ocular hypotony. Even though the IOP was not high, the intraluminal stents were completely removed in all cases 4 weeks after surgery when flow restriction by the stent for the prevention of postoperative ocular hypotony was no longer needed [4]. IOP was successfully controlled (8 to 17 mmHg) for 18, 34, and 36 months, respectively. There were no complications, such as postoperative ocular hypotony and tube-related complications (conjunctival erosion, tube exposure, migration, infection, and occlusion). These findings implied that a smaller tube may induce an effective IOP reduction with a low chance of complications which are in line with previous study results [3,4]. Given that aqueous drainage through the tube is mainly determined by the tube diameter, a modified BGI with a Finetube may have a lower chance of postoperative ocular hypotony compared to conventional BGI. An intraluminal stent can control the amount of aqueous drainage through the tube lumen after surgery—the stent can be retracted to increase drainage if necessary. Therefore, a safe, predictKorean J Ophthalmol 2017;31(1):90-91 ht tps://doi.org/10.3341/k jo.2017.31.1.90
منابع مشابه
Endophthalmitis associated with Glaucoma Shunt Intraluminal Stent Exposure.
Endophthalmitis post glaucoma drainage implant (GDI) surgery is rare, often associated with tube or plate exposure. We report a case of endophthalmitis following glaucoma shunt intraluminal stent exposure in a patient who underwent Baerveldt glaucoma implant surgery. Endophthalmitis following manipulation of intraluminal stents is a rare complication of GDIs but potentially vision threatening c...
متن کاملVitrectomy and Pars Plana Baerveldt Implant for Neovascular Glaucoma
APRIL 2011 I RETINA TODAY I 39 N eovascular glaucoma (NVG) is often caused by posterior segment disease, primarily diabetic retinopathy, central retinal vein occlusion, or carotid occlusive disease. Surgical options in NVG include trabeculectomy, trabeculectomy with mitomycin-C, cyclodestruction, and tube stent implants. An anterior segment approach to stent placement in patients with NVG may h...
متن کاملA case of WAGR syndrome in association with developmental glaucoma requiring bilateral Baerveldt glaucoma implants and subsequent tube repositioning
Glaucoma drainage device implantation is efficacious for the treatment of pediatric glaucoma patients when multiple angle surgeries fail. However, tube touching of the corneal endothelium is one of the major postoperative complications to deal with. A 15-month-old male patient with Wilms' tumor, aniridia, genitourinary anomalies, and mental retardation (WAGR) syndrome was diagnosed with bilater...
متن کاملComplications of Baerveldt glaucoma drainage implants.
OBJECTIVES To report the incidence and identify risk factors of postoperative complications after Baerveldt glaucoma drainage implantation. METHODS A retrospective review of the medical records of all patients at the Bascom Palmer Eye Institute, Miami, Fla, who underwent placement of a Baerveldt glaucoma drainage implant from October 1, 1992, through October 31, 1996, to determine demographic...
متن کاملSurgical Outcomes of Ahmed or Baerveldt Tube Shunt Implantation for medically Uncontrolled Traumatic Glaucoma
AIM To describe postoperative surgical success of either Ahmed or Baerveldt tube shunt implantation for eyes with medically uncontrolled traumatic glaucoma. MATERIALS AND METHODS A review was carried out to identify patients with traumatic glaucoma that required tube shunt implantation between 2009 and 2015 at Rutgers University in Newark, New Jersey, USA. Seventeen eyes from 17 patients met ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 31 شماره
صفحات -