The VeinViewer vascular imaging system worsens first-attempt cannulation rate for experienced nurses in infants and children with anticipated difficult intravenous access.

نویسندگان

  • Peter Szmuk
  • Jeffrey Steiner
  • Radu B Pop
  • Alan Farrow-Gillespie
  • Edward J Mascha
  • Daniel I Sessler
چکیده

BACKGROUND The VeinViewer (Luminetx, Memphis, TN) helps identify veins by projecting an image of subcutaneous vasculature on the skin surface. We tested the primary hypothesis that VeinViewer use improves cannulation success by skilled nurses in pediatric patients with anticipated difficult IV access. A secondary goal was to evaluate the relationship between obesity and cannulation success. METHODS Patients aged 0 to 18 years were included. Anticipated cannulation difficulty was evaluated with the difficult IV access score. All cannulations were performed by members of the Intravenous Access Team. Patients were randomized to: (1) routine IV catheter insertion; or (2) insertion facilitated by the VeinViewer. The primary outcome was first-attempt insertion success. The proportion of successful insertions was evaluated using Cochran-Mantel-Haenszel χ(2) analysis to adjust for any imbalanced baseline variables. The effect of obesity on cannulation success was evaluated with multivariable logistic regression. RESULTS Two hundred ninety-nine patients (49%) were randomly assigned to VeinViewer and 301 (51%) to routine cannulation. First-attempt cannulation success was 47% in patients assigned to VeinViewer vs 62% in patients assigned to routine cannulation, with an adjusted relative "risk" (95% confidence interval), of 0.76 (0.63-0.91). The Z-statistic of -3.6 crossed the "harm" boundary (Z < -2.41), with corresponding P value of 0.0003. The trial was stopped on statistical grounds since the harm boundary for the primary outcome was crossed. There was no association between first-attempt success and the 4-level categorization of obesity after adjusted for baseline variables (P = 0.94). CONCLUSIONS The VeinViewer worsened first-attempt IV insertion success by skilled nurses. Surprisingly, first-attempt success for IV cannulation was not worsened by obesity.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Efficacy of a near-infrared light device in pediatric intravenous cannulation: a randomized controlled trial.

OBJECTIVES To determine whether the use of a near-infrared light venipuncture aid (VeinViewer; Luminetx Corporation, Memphis, Tenn) would improve the rate of successful first-attempt placement of intravenous (IV) catheters in a high-volume pediatric emergency department (ED). METHODS Patients younger than 20 years with standard clinical indications for IV access were randomized to have IV pla...

متن کامل

Near-infrared imaging in intravenous cannulation in children: a cluster randomized clinical trial.

OBJECTIVE Intravenous cannulation is a widespread medical procedure that can be difficult in children. Visualization of veins with near-infrared (NIR) light might support intravenous cannulation. Therefore, we investigated the effectiveness of an NIR vascular imaging system (VascuLuminator) in facilitating intravenous cannulation in children in the operating room. METHODS This was a pragmatic...

متن کامل

Development of the A-DIVA Scale:

Placement of a peripheral intravenous catheter is a routine procedure in clinical practice, but failure of intravenous cannulation regularly occurs. An accurate and reliable predictive scale for difficult venous access creates the possibility to use other techniques in an earlier time frame. We aimed to develop a predictive scale to identify adult patients with a difficult intravenous access pr...

متن کامل

Ultrasound visual image-guided vs Doppler auditory-assisted radial artery cannulation in infants and small children by non-expert anaesthesiologists: a randomized prospective study.

BACKGROUND Cannulation of the radial artery in infants and small children can be challenging, even for the most experienced providers. Utilizing Doppler to aid in radial artery cannulation has been well described. Recent studies have demonstrated the efficacy of ultrasound (US) image-guided vascular access techniques in the paediatric population. The utility of these two techniques, when used b...

متن کامل

Comparing peripheral venous access between obese and normal weight children.

INTRODUCTION Intravenous (i.v.) access is sometimes a difficult, time-consuming, and highly frustrating procedure. Obesity is widely believed to be associated with difficult peripheral intravenous access (PIV) placement. This study examined the relationship between body mass index (BMI) and ease of venous access in children undergoing noncardiac surgical procedures. METHODS We prospectively c...