Implantation of cardioverter defibrillators without induction of ventricular fibrillation.

نویسنده

  • C D Swerdlow
چکیده

BACKGROUND The upper limit of vulnerability (ULV) is the weakest shock at which ventricular fibrillation (VF) is not induced by a T-wave shock. This study tested the hypothesis that a vulnerability safety margin based on the ULV can be used as an implantable cardioverter-defibrillator implantation criterion. METHODS AND RESULTS Implantable cardioverter-defibrillators were implanted in 80 patients if T-wave shocks did not induce VF and the baseline-rhythm R wave was >/=7 mV. The T-wave shock was 10 J in the first 45 patients (group A) and 15 J in the last 35 patients (group B). After inductionless implantations, the first VF shock was programmed to the T-wave shock plus 5 J. If T-wave shocks induced VF, the ULV was measured and the first shock was programmed to the ULV+5 J. Inductionless implantations were performed in 58 patients (72%), 28 in group A (62%) and 30 in group B (86%; P=0.04). If T-wave scanning had been done at 15 J in group A patients, inductionless implantations could have been performed in 84% of them. At 3 months, VF was induced twice during electrophysiological study in 75 patients (94%). All VFs were detected in 80% of implantable cardioverter-defibrillator recipients using a vulnerability safety margin based on a T-wave scan at 15.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

Atrial fibrillation coexisting with ventricular tachycardia: a challenge for dual chamber defibrillators.

Atrial fibrillation is a very common arrhythmia in patients who need an implantable cardioverter-defibrillator (ICD) because of life threatening ventricular tachyarrhythmias. The prevalence of atrial fibrillation at the time of implantation has been calculated to be as high as 20% and it has been reported that during the lifespan of the ICD more than 50% of patients may develop atrial fibrillat...

متن کامل

Success of Device-Based Methods for Atrial Tachyarrhythmia Induction during Implantation of an Implantable Cardioverter Defibrillator with Atrial Therapy

In addition to standard ventricular therapy, cardioversion using an implantable cardioverter defibrillator (ICD) with atrial therapy is one of the many new options available for treating atrial tachyarrhythmias. New data suggest that shock efficacy for AF depends on identifying atrial defibrillation thresholds (DFT) and programming the first atrial shock to at least two times the atrial DFT [1]...

متن کامل

Death Without Prior Appropriate Implantable Cardioverter-Defibrillator Therapy

Background—Implantable cardioverter-defibrillators (ICDs) improve survival in selected patients with left ventricular systolic dysfunction in randomized trials. Competing death without prior appropriate ICD therapy might preclude benefit from ICD implantation in a less selected routine-care population. Methods and Results—We selected all patients with ischemic or dilated cardiomyopathy with an ...

متن کامل

Death Without Prior Appropriate Implantable Cardioverter-Defibrillator Therapy A Competing Risk Study

Background—Implantable cardioverter-defibrillators (ICDs) improve survival in selected patients with left ventricular systolic dysfunction in randomized trials. Competing death without prior appropriate ICD therapy might preclude benefit from ICD implantation in a less selected routine-care population. Methods and Results—We selected all patients with ischemic or dilated cardiomyopathy with an ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Circulation

دوره 103 17  شماره 

صفحات  -

تاریخ انتشار 2001