Quality assurance methodology for Varian RapidArc treatment plans

نویسندگان

  • Ileana Iftimia
  • Eileen T. Cirino
  • Li Xiong
  • Herbert W. Mower
چکیده

With the commercial introduction of the Varian RapidArc, a new modality for treatment planning and delivery, the need has arisen for consistent and efficient techniques for performing patient-specific quality assurance (QA) tests. In this paper we present our methodology for a RapidArc treatment plan QA procedure. For our measurements we used a 2D diode array (MapCHECK) embedded at 5 cm water equivalent depth in MapPHAN 5 phantom and an Exradin A16 ion chamber placed in six different positions in a cylindrical homogeneous phantom (QUASAR). We also checked the MUs for the RapidArc plans by using independent software (RadCalc). The agreement between Eclipse calculations and MapCHECK/MapPHAN5 measurements was evaluated using both absolute distance-to-agreement (DTA) and gamma index with 10% dose threshold (TH), 3% dose difference (DD), and 3 mm DTA. The average agreement was 94.4% for the DTA approach and 96.3% for the gamma index approach. In high-dose areas, the discrepancy between calculations and ion chamber measurements using the QUASAR phantom was within 4.5% for prostate cases. For the RadCalc calculations, we used the average SSD along the arc; however, for some patients the agreement for the MUs obtained with RadCalc versus Eclipse was inadequate (discrepancy > 5%). In these cases, the plan was divided into partial arc plans so that RadCalc could perform a better estimation of the MUs. The discrepancy was further reduced to within ~4% using this approach. Regardless of the variation in prescribed dose and location of the treated areas, we obtained very good results for all patients studied in this paper.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of RapidArc and Static Gantry Intensity-modulated Radiotherapy for Nasopharyngeal Carcinoma

Objective: RapidArc (Varian Medical Systems, Palo Alto, CA) is a new volumetric-modulated arc radiotherapy technique with continuous modulation of gantry speed, dose rate, and multileaf collimator shape. Taking advantages of dose painting capabilities along with rotational delivery, RapidArc has shown promise in the treatment of a wide range of tumours, including intracranial, head and neck, pr...

متن کامل

Comparison of four commercial devices for RapidArc and sliding window IMRT QA

For intensity-modulated radiation therapy, evaluation of the measured dose against the treatment planning calculated dose is essential in the context of patient-specific quality assurance. The complexity of volumetric arc radiotherapy delivery attributed to its dynamic and synchronization nature require new methods and potentially new tools for the quality assurance of such techniques. In the p...

متن کامل

Teflon cylindrical phantom for delivery quality assurance of SBRT

At our institution the standard delivery quality assurance (DQA) procedure for tomotherapy plans is accomplished with a water equivalent phantom, EDR2 film, and ion chamber point-dose measurements. Most plans deliver at most 5 Gy to the dose plane; however, recently a stereotactic body radiotherapy (SBRT) protocol has produced plans delivering upwards of 12 Gy to the film plane. EDR2 film satur...

متن کامل

Commissioning and early experience with a new-generation low-energy linear accelerator with advanced delivery and imaging functionalities

BACKGROUND A new-generation low-energy linear accelerator (UNIQUE) was introduced in the clinical arena during 2009 by Varian Medical Systems. The world's first UNIQUE was installed at Oncology Institute of Southern Switzerland and put into clinical operation in June 2010. The aim of the present contribution was to report experience about its commissioning and first year results from clinical o...

متن کامل

Is the dose distribution distorted in IMRT and RapidArc treatment when patient plans are swapped across beam‐matched machines?

The purpose of this study is to evaluate the degree of dose distribution distortion in advanced treatments like IMRT and RapidArc when patient plans are swapped across dosimetrically equivalent so-called "beam-matched" machines. For this purpose the entire work is divided into two stages. At forefront stage all basic beam properties of 6 MV X-rays like PDD, profiles, output factors, TPR20/10 an...

متن کامل