Downloadable algorithm to reduce inappropriate shocks caused by fractures of implantable cardioverter-defibrillator leads.
نویسندگان
چکیده
BACKGROUND The primary method for monitoring implantable cardioverter-defibrillator lead integrity is periodic measurement of impedance. Sprint Fidelis leads are prone to pace-sense lead fractures, which commonly present as inappropriate shocks caused by oversensing. METHODS AND RESULTS We developed and tested an algorithm to enhance early identification of lead fractures and to reduce inappropriate shocks. This lead-integrity algorithm, which can be downloaded into presently implanted implantable cardioverter-defibrillators, alerts the patient and/or physician when triggered by either oversensing or excessive increases in impedance. To reduce inappropriate shocks, the lead-integrity algorithm increases the number of intervals to detect (NID) ventricular fibrillation when triggered. The lead-integrity algorithm was tested on data from 15 970 patients with Fidelis leads (including 121 with clinically diagnosed fractures) and 95 other fractured leads confirmed by analysis of returned product. The effect of the NID on inappropriate shocks was tested in 92 patients with 927 shocks caused by lead fracture. Increasing the NID reduced inappropriate shocks (P<0.0001). The lead-integrity algorithm provided at least a 3-day warning of inappropriate shocks in 76% (95% CI, 66 to 84) of patients versus 55% (95% CI, 43 to 64) for optimal impedance monitoring (P=0.007). Its positive predictive value was 72% for lead fractures and 81% for lead fractures or header-connector problems requiring surgical intervention. The false-positive rate was 1 per 372 patient-years of monitoring. CONCLUSIONS A lead-integrity algorithm developed for download into existing implantable cardioverter-defibrillators increases short-term warning of inappropriate shocks in patients with lead fractures and reduces the likelihood of inappropriate shocks. It is the first downloadable RAMware to enhance the performance of nominally functioning implantable cardioverter-defibrillators and the first implantable cardioverter-defibrillator monitoring feature that triggers real-time changes in ventricular fibrillation detection parameters to reduce inappropriate shocks.
منابع مشابه
Arrhythmia/Electrophysiology Downloadable Software Algorithm Reduces Inappropriate Shocks Caused by Implantable Cardioverter-Defibrillator Lead Fractures A Prospective Study
Background—Downloadable software upgrades are common in consumer electronics but not in implantable medical devices. Fractures of implantable cardioverter-defibrillator (ICD) leads present commonly as inappropriate shocks. A lead-integrity alert (LIA) designed to reduce inappropriate shocks is the first software download approved to enhance nominally functioning, previously implanted ICDs. Meth...
متن کاملDownloadable software algorithm reduces inappropriate shocks caused by implantable cardioverter-defibrillator lead fractures: a prospective study.
BACKGROUND Downloadable software upgrades are common in consumer electronics but not in implantable medical devices. Fractures of implantable cardioverter-defibrillator (ICD) leads present commonly as inappropriate shocks. A lead-integrity alert (LIA) designed to reduce inappropriate shocks is the first software download approved to enhance nominally functioning, previously implanted ICDs. ME...
متن کاملA better method for preventing adverse clinical events caused by implantable cardioverter-defibrillator lead fractures?
In October 2007, the Medtronic Sprint Fidelis implantable cardioverter-defibrillator (ICD) lead was voluntarily withdrawn from the market by the manufacturer because of a higher-than-expected fracture rate.1 Most Sprint Fidelis fractures involve the pace-sense component of the lead, which may cause oversensing of electrical artifacts, undersensing of ventricular depolarizations, and/or loss of ...
متن کاملArrhythmia/Electrophysiology Longevity of Sprint Fidelis Implantable Cardioverter- Defibrillator Leads and Risk Factors for Failure Implications for Patient Management
Background—Sprint Fidelis (Fidelis) implantable cardioverter-defibrillator leads are prone to fractures that have caused adverse events, primarily inappropriate shocks, and a few reported deaths. More than 100 000 patients have Fidelis leads. No independent multicenter long-term performance information exists for this lead, and single-center studies suggest that certain patients are at increase...
متن کاملLongevity of Sprint Fidelis implantable cardioverter-defibrillator leads and risk factors for failure: implications for patient management.
BACKGROUND Sprint Fidelis (Fidelis) implantable cardioverter-defibrillator leads are prone to fractures that have caused adverse events, primarily inappropriate shocks, and a few reported deaths. More than 100 000 patients have Fidelis leads. No independent multicenter long-term performance information exists for this lead, and single-center studies suggest that certain patients are at increase...