Bacterial presence on flexible endoscopes vs time since disinfection
نویسندگان
چکیده
AIM To correlate the length of endoscope hang time and number of bacteria cultured prior to use. METHODS Prospectively, we cultured specimens from 19 gastroscopes, 24 colonoscopes and 5 side viewing duodenoscopes during the period of 2011 to 2015. A total of 164 results had complete data denoting date of cleansing, number of days stored and culture results. All scopes underwent initial cleaning in the endoscopy suite utilizing tap water, and then manually cleaned and flushed. High level disinfection was achieved with a Medivator© DSD (Medivator Inc., United States) automated endoscope reprocessor following manufacturer instructions, with Glutacide® (Pharmax Limited, Canada), a 2% glutaraldehyde solution. After disinfection, all scopes were stored in dust free, unfiltered commercial cabinets for up to 7 d. Prior to use, all scopes were sampled and plated on sheep blood agar for 48 h; the colony count was obtained from each plate. The length of endoscope hang time and bacterial load was analyzed utilizing unpaired t-tests. The overall percentage of positive and negative cultures for each type of endoscope was also calculated. RESULTS All culture results were within the acceptable range (less than 200 cfu/mL). One colonoscope cultured 80 cfu/mL after hanging for 1 d, which was the highest count. ERCP scopes cultured at most 10 cfu, this occurred after 2 and 7 d, and gastroscopes cultured 50 cfu/mL at most, at 1 d. Most cultures were negative for growth, irrespective of the length of hang time. Furthermore, all scopes, with the exception of one colonoscope which had two positive cultures (each of 10 cfu/mL), had at most one positive culture. There was no significant difference in the number of bacteria cultured after 1 d compared to 7 d when all scopes were combined (day 2: P = 0.515; day 3: P = identical; day 4: P = 0.071; day 5: P = 0.470; day 6: P = 0.584; day 7: P = 0.575). There was also no significant difference in the number of bacteria cultured after 1 day compared to 7 d for gastroscopes (day 2: P = 0.895; day 3: P = identical; day 4: P = identical; day 5: P = 0.893; day 6: P = identical; day 7: P = 0.756), colonoscopes (day 2: P = 0.489; day 4: P = 0.493; day 5: P = 0.324; day 6: P = 0.526; day 7: P = identical), or ERCP scopes (day 2: P = identical; day 7: P = 0.685). CONCLUSION There is no correlation between hang time and bacterial load. Endoscopes do not need to be reprocessed if reused within a period of 7 d.
منابع مشابه
Reprocessing endoscopes: United States perspective.
Endoscopes are used frequently for the diagnosis and therapy of medical disorders. For example, greater than 10000000 gastrointestinal endoscopic procedures are performed each year in the United States. Failure to employ appropriate cleaning and disinfection/sterilization of endoscopes has been responsible for multiple nosocomial outbreaks and serious, sometimes life-threatening, infections. Fl...
متن کاملMicrobiologic profile of flexible endoscope disinfection in two Brazilian hospitals.
BACKGROUND [corrected] Endoscopes are routinely used in hospitals and clinics of the world and they can be potential sources of cross-infection when the decontamination process is unsuitable AIM The routines of flexible endoscope (bronchoscopes, esophagogastroduodenoscopes and colonoscopes) disinfection procedures used in two Brazilian university hospitals were evaluated during a 3-year perio...
متن کاملJoint AUA/SUNA white paper on reprocessing of flexible cystoscopes.
PURPOSE Numerous federal bodies and professional societies have produced guidelines and standards for the reprocessing of endoscopes, but few specifically address the reprocessing of cystoscopes. MATERIALS AND METHODS This document presents a summary of the current recommendations for reprocessing flexible cystoscopes, and highlights particular aspects of instrument reprocessing that are uniq...
متن کاملATP measurement as method to monitor the quality of reprocessing flexible endoscopes
Insufficient performance of cleaning and disinfection of flexible endoscopes can pose an infection risk to patients. Actually quality of reprocessing is checked by performing microbiological cultures. Unfortunately, their results are not available on the same day so that more rapid methods are desirable. We compared the ATP (adenosine triphosphate) bioluminescence for hygiene checking of the re...
متن کاملModeling microbial survival in buildup biofilm for complex medical devices
BACKGROUND Flexible endoscopes undergo repeated rounds of patient-use and reprocessing. Some evidence indicates that there is an accumulation or build-up of organic material that occurs over time in endoscope channels. This "buildup biofilm" (BBF) develops as a result of cyclical exposure to wet and dry phases during usage and reprocessing. This study investigated whether the BBF matrix represe...