Noise correlation matrix for multiple RF coils derived from first principles of statistical physics

نویسندگان

  • Y. Wang
  • R. Brown
  • P. Spincemaille
  • R. Lee
چکیده

Y. Wang, R. Brown, P. Spincemaille, R. Lee Radiology, Weill Medical College of Cornell University, New York, NY, United States, Radiology, New York University, New York, NY, United States Introduction Noise correlation has been a controversial and confusing topic in MRI literature (1,2), due to the lack of a clear definition of thermal noise and the complexity of circuitry in MR scanners. The formulation outlined in the original Nyquist paper does not provide a straightforward account for noise correlation among two circuits (3). With the widespread use of multiple rf channel technology on MR scanners, the noise correlation matrix among coil elements is becoming a very important issue for coil design (4,5), and it is desired to have a fundamental understanding of the noise correlation. Here we attempt to elucidate the noise correlation from the first principle of thermal statistical physics. Theory The energy stored in a system of circuits is (6), H(q1...qn, i1...in) = ∑abLabiaib/2 + ∑aqa /2Ca, where ia is the current in coil a, qa the charge, Ca the capacitance, and Lab the mutual inductance between coils a and b. The thermal probability for various current values is determined by the Boltzman distribution. The noise correlation matrix is derived from the thermal average of current fluctuation (7): = ∫iaibexp(-∑cdLcdicid/2kT)di1...din/ ∫exp(-∑cdLcdicid/2kT)di1...din = kT(L )ab. Hence, Ψab ≡ = kT(L)ab. [1] Eq.1 characterizes the total noise correlation. Noise correlation at a given frequency is derived from the fluctuation-dissipation theorem with E/iω = f the “force” for current and iω/Z = α(ω) the generalized susceptibility (7,8): Ψab ≡ = 4kT Re((Z)ab). [2] Here Z is the coil impedance matrix. The relation between the noise correlation over all frequencies and that at one frequency is Ψab = ∫0 dω/2π Ψab, as derived from the Kramers-Kronig relation (7), α’(0)− α’(∞)=2/π ∫0dω α”(ω)/ω, where α’ and α” are the real and imaginary part of α. From the definition of impedance matrix, α’(0) = 0, α’(∞) = -L, α”(ω)/ω = Re((Z)). Therefore, (L)ab = 2/π ∫0dω Re((Z)ab). [3] The above theory is examplified in the case of a single coil made of R, L and C in series. <(i)> = 4kT R/(R+(ωL-1/ωC)), = 4kT ∫0 dω/2π R/(R+(ωL-1/ωC)) = kT/L (9). Materials and methods Two circular coils with 13.2 cm diameter were constructed. Experiments were performed on this coil system over a large tank of water. The spacing between the two coils was variable to allow a range of mutual inductance. The impedance matrix at various frequencies was measured using a HP network analyzer. Measurements of noise (voltage converted to current using the impedance matrix) were performed on a 1.5T GE SIGNA CV/i MR system. The RF excitation was disabled during imaging for noise collection. The mutual inductance matrix was estimated according to method by Grover (10). Results The network analyzer measured total noise (right side of Eq.3, Area(Re((Z)12)))was linearly correlated to the estimated inverse inductance (left side of Eq.3, (L)12) for the two coil system (Fig.1a), r = .95. The MR scanner measured noise matrix was linearly correlated to real part of the inverse of the coil impedance matrix (Fig.1b), r = .94. Discussion These preliminary experiments confirm the theoretical predictions on noise correlation (Eqs.2&3). Small errors in the experimental measurements may be attributed to weighting by the coil resonance spectrum (Q values) and effects from amplifiers. The theory presented here clarifies confusions in MRI literature. Eq.1 indicates that the total noise correlation is zero when there is no mutual inductance, consistent with the thermodynamic argument in Ref. 1. However, the noise correlation at a given resonance frequency is determined by the impedance matrix (Eq.2), as suggested in Ref. 2. Eqs.1&2 are consistent, as indicated by the KramersKronig relation Eq.3. REFERENCES 1. Jesmanowicz A et al, MRM 20:36-47 (1991). 2. Redpath TW, MRM 24: 85-9 (1992); British J Radiology 71:704-7 (1998). 3. Nyquist H, Phys Rev 32:110-3 (1928). 4. Weiger M, et al, MRM 45:495-504 (2001). 5. Ohliger MA, et al, MRM 52:628-39 (2004). 6. Landau and Lifshitz, Electrodynamics of Continuous Media, 2002. 7. Landau and Lifshitz, Statistical Physics, Part I&II;, 2003. 8. Callen HB, et al, Phys Rev, 83:34-40 (1951). 9. Feynman’s Lecture on Physics, v.1, 1977. 10. Grover FW, Inductance calculations, working formulas and tables, 1946.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences

Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of  non-homogeneity in RF receive and transmit. This can be the most effective source of error in  quantitative  studies  in  MRI  imaging.  Part  of  this  non-homogeneity  demonstrates  the  characteristics of RF coil and part of it is due to the interaction of RF field with the material being  imaged...

متن کامل

Effects of RF-EMF Exposure from GSM Mobile Phones on Proliferation Rate of Human Adipose-derived Stem Cells: An In-vitro Study

Background:As the use of mobile phones is increasing, public concern about the harmful effects of radiation emitted by these devices is also growing. In addition, protection questions and biological effects are among growing concerns which have remained largely unanswered. Stem cells are useful models to assess the effects of radiofrequency electromagnetic fields (RF-EMF) on other cell lines. S...

متن کامل

Radiative principles for efficient RF transmission at UHF : radiative antennas and travelling waves

RF coils can be classified in RF receive coils and RF transmit coils. RF receive coil are designed to create a maximal sensitivity to the NMR signals emitted by the spins while picking up a minimal amount of thermal noise. The function of RF transmit coils is to generate a homogeneous B1+ field in the imaging region while inducing minimal RF power deposition that results in undesired RF tissue ...

متن کامل

In vivo multiple-mouse MRI at 7 Tesla.

We developed a live high-field multiple-mouse magnetic resonance imaging method to increase the throughput of imaging studies involving large numbers of mice. Phantom experiments were performed in 7 shielded radiofrequency (RF) coils for concurrent imaging on a 7 Tesla MRI scanner outfitted with multiple transmit and receive channels to confirm uniform signal-to-noise ratio and minimal ghost ar...

متن کامل

Non-uniformity of Clinical Head, Head and Neck, and Body Coils in Magnetic Resonance Imaging (MRI)

Introduction Signal intensity uniformity in a magnetic resonance (MR) image indicates how well the MR imaging (MRI) system represents an object. One of the major sources of image non-uniformity in high-field MRI scanners is inhomogeneity of radio-frequency coil. The aim of this study was to investigate non-uniformity in head, head and neck, and body coils and compare the obtained results to det...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2005