International collaborative study for the calibration of a proposed reference preparation for thromboplastin, human recombinant, plain. On behalf of the Subcommittee on Control of Anticoagulation.

نویسندگان

  • A Tripodi
  • V Chantarangkul
  • B Negri
  • M Clerici
  • P M Mannucci
چکیده

Stocks of the International Reference Preparation (IRP) for thromboplastin, human, plain, coded BCT/253 and held by the World Health Organization (WHO) are nearly exhausted and must be replaced. For practical reasons the choice of the replacement candidate was restricted to two available human recombinant preparations which were coded as X/95 and Y/95 and calibrated in an international collaborative study involving 19 laboratories from Europe, Australia, Canada and Argentina. To minimize the differences between routes of calibration, the two candidates were calibrated against the existing WHO-IRP from human, rabbit and bovine origin and the final ISI was the resultant average value. On the basis of predefined criteria (i.e., within- and between-laboratory precision of the calibration and the conformity to the calibration model), X/95 was the preferred candidate. The assigned ISI (SE of the mean) value is 0.940 (0.0060) and the interlaboratory coefficient of variation 4.7%.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Noninvasive prenatal diagnosis of a case of Down syndrome due to robertsonian translocation by massively parallel sequencing of maternal plasma DNA.

1. Remijn JA, Lucas S, Wildeboer B, van Suijlen JDE, Adriaansen HJ. Strongly increased international normalized ratio with recombinant Neoplastin R compared with tissue extract Neoplastin Plus in patients initiating oral anticoagulation therapy: implications for anticoagulation dosage. Clin Chem 2008;54:1929–31. 2. Roussi J, Drouet L, Samama M, Sié P. French multicentric evaluation of recombina...

متن کامل

Determination of Biological Activity of Recombinant Reteplase Using Clot Lysis Time and Activated Partial Thromboplastin Time (APTT) Lysis Methods: A Comparative Study

Recombinant plasminogen activator (reteplase) is a third generation thrombolytic agent which has been used on coronary artery thrombosis and acute myocardial infarction. Clot lysis assay is usually considered as a unique method to evaluate biological activity of reteplase. In this study biological activity of reteplase was determined by APTT (activated partial thromboplastin time) lysis method....

متن کامل

Determination of Biological Activity of Recombinant Reteplase Using Clot Lysis Time and Activated Partial Thromboplastin Time (APTT) Lysis Methods: A Comparative Study

Recombinant plasminogen activator (reteplase) is a third generation thrombolytic agent which has been used on coronary artery thrombosis and acute myocardial infarction. Clot lysis assay is usually considered as a unique method to evaluate biological activity of reteplase. In this study biological activity of reteplase was determined by APTT (activated partial thromboplastin time) lysis method....

متن کامل

European Concerted Action on Anticoagulation. A multicentre calibration study of WHO international reference preparations for thromboplastin, rabbit (RBT/90) and human (rTF/95).

A 10 centre calibration was performed after six years to determine the international sensitivity index (ISI) of rTF/95 relative to RBT/90, and to assess any international normalised ratio (INR) bias compared with the original multicentre calibration. After exclusion of one outlying centre, the follow up calibration gave a mean ISI for rTF/95 of 0.99, which although a small difference, is signif...

متن کامل

Multicentre ISI assignment and calibration of the INR measuring range of a new point-of-care system designed for home monitoring of oral anticoagulation therapy.

The new CoaguChek XS system is designed for use in patient selftesting. It is the successor of the current CoaguChek S system. The detection principle is based on the amperometric measurement of the thrombin activity initiated by starting the coagulation cascade using a human recombinant thromboplastin. This study was performed to assign the International SEnsitivity Index (ISI) to the new test...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Thrombosis and haemostasis

دوره 79 2  شماره 

صفحات  -

تاریخ انتشار 1998