Oblique reconstructions in tomosynthesis. II. Super-resolution.

نویسندگان

  • Raymond J Acciavatti
  • Andrew D A Maidment
چکیده

PURPOSE In tomosynthesis, super-resolution has been demonstrated using reconstruction planes parallel to the detector. Super-resolution allows for subpixel resolution relative to the detector. The purpose of this work is to develop an analytical model that generalizes super-resolution to oblique reconstruction planes. METHODS In a digital tomosynthesis system, a sinusoidal test object is modeled along oblique angles (i.e., "pitches") relative to the plane of the detector in a 3D divergent-beam acquisition geometry. To investigate the potential for super-resolution, the input frequency is specified to be greater than the alias frequency of the detector. Reconstructions are evaluated in an oblique plane along the extent of the object using simple backprojection (SBP) and filtered backprojection (FBP). By comparing the amplitude of the reconstruction against the attenuation coefficient of the object at various frequencies, the modulation transfer function (MTF) is calculated to determine whether modulation is within detectable limits for super-resolution. For experimental validation of super-resolution, a goniometry stand was used to orient a bar pattern phantom along various pitches relative to the breast support in a commercial digital breast tomosynthesis system. RESULTS Using theoretical modeling, it is shown that a single projection image cannot resolve a sine input whose frequency exceeds the detector alias frequency. The high frequency input is correctly visualized in SBP or FBP reconstruction using a slice along the pitch of the object. The Fourier transform of this reconstructed slice is maximized at the input frequency as proof that the object is resolved. Consistent with the theoretical results, experimental images of a bar pattern phantom showed super-resolution in oblique reconstructions. At various pitches, the highest frequency with detectable modulation was determined by visual inspection of the bar patterns. The dependency of the highest detectable frequency on pitch followed the same trend as the analytical model. It was demonstrated that super-resolution is not achievable if the pitch of the object approaches 90°, corresponding to the case in which the test frequency is perpendicular to the breast support. Only low frequency objects are detectable at pitches close to 90°. CONCLUSIONS This work provides a platform for investigating super-resolution in oblique reconstructions for tomosynthesis. In breast imaging, this study should have applications in visualizing microcalcifications and other subtle signs of cancer.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Observation of super-resolution in digital breast tomosynthesis.

PURPOSE Digital breast tomosynthesis (DBT) is a 3D x-ray imaging modality in which tomographic sections of the breast are generated from a limited range of tube angles. Because oblique x-ray incidence shifts the image of an object in subpixel detector element increments with each increasing projection angle, it is demonstrated that DBT is capable of super-resolution (i.e., subpixel resolution)....

متن کامل

Proposing an Acquisition Geometry That Optimizes Super-Resolution in Digital Breast Tomosynthesis

In digital breast tomosynthesis (DBT), oblique x-ray incidence shifts the image of an object in sub-pixel detector element increments with each projection angle. Our previous work has shown that DBT is capable of superresolution as a result of this property. Although super-resolution is achievable over a broad range of positions for frequencies parallel to the chest wall side of the breast supp...

متن کامل

Investigating the Potential for Super-Resolution in Digital Breast Tomosynthesis

Digital breast tomosynthesis (DBT) is an emerging 3D x-ray imaging modality in which tomographic sections of the breast are generated from a limited range of tube angles. Because non-normal x-ray incidence causes the image of an object to be translated in sub-pixel increments with increasing projection angle, it is demonstrated in this work that DBT is capable of super-resolution (i.e., sub-pix...

متن کامل

Dynamic Contrast-Enhanced Digital Breast Tomosynthesis

PURPOSE The purpose of this work is to characterize the spatial dependence of the in-plane 35 modulation transfer function (MTF) in tomosynthesis and investigate the influence of the acquisition geometry. METHODS The in-plane MTF was calculated from measured edge spread functions using a custom 40 test tool consisting of a 0.250 mm thick lead sheet affixed to a 5.88 mm acrylic sheet. Images wer...

متن کامل

Calculation of OTF, NPS, and DQE for Oblique X-Ray Incidence on Turbid Granular Phosphors

Digital breast tomosynthesis (DBT) is an imaging modality in which tomographic sections of the breast are generated from a limited range of x-ray tube angles. One drawback of DBT is resolution loss in the oblique projection images. The purpose of this work is to extend Swank’s formulation of the transfer functions of turbid granular phosphors to oblique x-ray incidence, using the diffusion appr...

متن کامل