Evaluation of pressure support ventilation with seven different ventilators using Active Servo Lung 5000.

نویسندگان

  • Michihisa Terado
  • Shingo Ichiba
  • Osamu Nagano
  • Yoshihito Ujike
چکیده

In modern emergency and critical care, physicians tend to choose the mode of mechanical ventilation based on spontaneous breathing for the purpose of promoting discharge of pulmonary secretion and preventing atelectasis in patients with acute respiratory insufficiency. However, we often observe "differences in recovery" among patients treated using the same PSV settings beyond "differences in individual characteristics." We evaluated the Pressure Support Ventilation (PSV) mode aiming to certify the difference among 7 representative mechanical ventilators using the Active Servo Lung 5000 (ASL5000) respiratory simulation system. The following parameters were measured: The time delay that resulted in the lowest inspiratory pressure from the point at which the ventilator recognized spontaneous breathing (TD), the lowest inspiratory airway pressure (cmH2O) generated prior to the initiation of PSV (DeltaPaw), the work of breathing while triggering required to achieve the lowest inspiratory negative pressure from the beginning of inspiratory support (WOBtrig), and the inspiratory work of breathing (WOBi). The mean TD of the Puritan-Bennett type 840 (PB840) was signifi cantly shorter than those of other ventilators (p0.01). The WOBtrig of the PB840 was significantly lower than those of others (p0.01). However, the WOBi values of the Servo-I and T-Bird were greater than the others, with the Evita series showing the smallest WOBi of the 7 ventilators tested. According to this simulation study using ASL 5000, we concluded that PB840 was the most rapid response ventilator, but the Evita series was the gentlest mechanical ventilator among 7 ventilators from the standpoint of the total work of breathing during the inspiration phase in the setting of PSV.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Leak compensation in positive pressure ventilators: a lung model study.

Leak compensating abilities of six different positive pressure ventilators commonly used to deliver noninvasive positive pressure ventilation, including the bilevel positive airway pressure (BiPAP) S/T-D and Quantum (Respironics Inc, Murrysville, PA, USA), 335 and O'NYX (Mallinckrodt Inc, St Louis, MO, USA), PLV 102 (Respironics), and Siemens Servo 900C (Siemens Inc, Danvers, MA, USA). Using a ...

متن کامل

Harms of unintentional leaks during volume targeted pressure support ventilation.

BACKGROUND Volume targeted pressure support ventilation (VT-PSV) is a hybrid mode increasingly used to maintain a minimal tidal volume (VT) by automatically adjusting the level of inspiratory pressure. The objective of the study was to determine the ability of home ventilators to maintain the preset minimal VT during unintentional leaks in a VT-PSV mode. METHODS Seven ventilators were tested ...

متن کامل

Bench study on active exhalation valve performance.

BACKGROUND Ventilator exhalation-valve performance during the expiratory phase has been studied in depth. An active exhalation valve uses servo-control technology that allows gas to be released from the exhalation valve during the inspiratory phase if the patient makes an expiratory effort. We conducted a bench study of active exhalation valve response to expiratory effort during the inspirator...

متن کامل

Work of breathing in adaptive pressure control continuous mandatory ventilation.

BACKGROUND Adaptive pressure control is a mode of mechanical ventilation where inflation pressure is adjusted by the ventilator to achieve a target tidal volume (VT). This means that as patient effort increases, inflation pressure is reduced, which may or may not be clinically appropriate. The purpose of this study was to evaluate the relationship between ventilator work output and patient effo...

متن کامل

An evaluation of home volume ventilators that support open-circuit mouthpiece ventilation.

BACKGROUND Open-circuit mouthpiece ventilation (MPV) is a form of noninvasive ventilation that can be used to provide portable daytime ventilatory support for neuromuscular patients with chronic respiratory failure. MPV has been reported to reduce the risk of respiratory infection due to tracheostomy, and to improve cough and voice function and patient quality of life. Despite these potential b...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Acta medica Okayama

دوره 62 2  شماره 

صفحات  -

تاریخ انتشار 2008