In vitro evaluation of the accuracy of five different electronic apex locators for determining the working length of endodontically retreated teeth.

نویسندگان

  • Aqeel Khalil Ebrahim
  • Reiko Wadachi
  • Hideaki Suda
چکیده

The aim of this study was to evaluate the accuracy of five electronic apex locators (EALs) in determining the working length (WL) of teeth after removal of the root canal obturation materials. A total of 32 extracted straight, single-rooted teeth were used. The actual canal length (AL) was determined and the WL was established by subtracting 0.5 mm from the AL. The root canals were instrumented and divided into two groups. One group (n = 6) served as control, while the other group (n = 26) was the experimental group. In the experimental group, the root canals were obturated using vertically compacted gutta-percha with AH 26 sealer. In both groups, the access cavities were restored with a provisional restoration and stored for 15 days at 37 degrees C and 100% humidity. The root canal obturation material was removed, and the teeth were then mounted in an experimental apparatus. Five EALs were used: Dentaport ZX, ProPex, Foramatron D10, Apex NRG and Apit 7. For the electronic measurement of canal length, a size 25 K-file was used. During measurement, the canal was irrigated with 2.5% sodium hypochlorite. The difference (D) between the AL and the electronically determined length (EDL), AL-EDL, was calculated and recorded for each measurement. Data were analysed by two-way anova and Fisher's PLSD test. In both groups, statistically significant differences were found among the EALs (P < 0.01). In conclusion, the Dentaport ZX, ProPex and Foramatron D10 were more accurate than the other two EALs in determining the WL in teeth after removal of the root canal obturation materials. However, the Apex NRG and Apit 7 were also reliable for determination of the WL in the majority of the cases.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination

In vitro evaluation of the accuracy of two new electronic apex locators in root canal length determination Dr. Hasheminia, S.M.* - Dr. Jafari, N.** *Associate Professor of Endodontics Dept., Faculty of Dentistry, Isfahan University of Medical Sciences. ** Dentist. Abstract Background & Aim: Radiography is still the common method of measuring working length in root canal therapy. While, gag refl...

متن کامل

ارزیابی دقت دو دستگاه آپکس یاب الکترونیکی (Electronic Apex Locator) از نسل چهارم در تعیین طول کانال (In-vivo)

Background and Aim: Determining correct working length in root canal therapy is of great importance in successful treatment. The purpose of this study was to compare the accuracy of two electronic apex locators of fourth generation, Raypex4 and Apit7, in determining working length in vivo. Materials and Methods: In this test evaluation study performed at Dental School of Tehran University, a to...

متن کامل

The effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants

BACKGROUND AND AIM: The aim of the present study was to evaluate the effect of different needle-insertion depths on the accuracy of the Root ZX II and Root ZX mini apex locators in the presence of various irrigants. METHODS: Ninety extracted single-canal human teeth were used in nine experimental groups. The coronal one-third of the canals was preflared, and the length to the major foramen was ...

متن کامل

Evaluation of the accuracy of two apex locators in endodontic treatment and retreatments: an ex vivo study

Introduction: Working length determination is important in successful endodontic treatment and retreatment.This study evaluated the accuracy of two electronic apex locators Root ZX and Raypex&reg;6 (EALs) in determining the electronic working length (EWL) of the root canals in endodontic treatment and retreatment. Materials and Methods: Access cavities were prepared on forty extracted, singl...

متن کامل

Comparison of Working Length Determination Using Apex Locator and Conventional Radiography in Curved Canals-an In Vitro Study

Introduction: The exact measurement of the root canal length is of great importance in root canal therapy. Although, determination of the root canal length through radiography is the most common method, recently electronic apex locators are being used to determine working length and decrease the number of radiographs. The purpose of this study was to compare the accuracy of an electronic apex l...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Australian endodontic journal : the journal of the Australian Society of Endodontology Inc

دوره 33 1  شماره 

صفحات  -

تاریخ انتشار 2007