Chemical shift MR imaging in the lumbar vertebra: the effect of field strength, scanner vendors and flip angles in repeatability of signal intensity index measurement
نویسندگان
چکیده
BACKGROUND To evaluate the reproducibility of signal intensity index (SII) measurements with MRI systems from different vendors and with different field strengths, and to test the effectiveness of flip angle. METHODS Thirty-two healthy volunteers (mean age 35.3 ± 9.3 years) were enrolled in this ethics committee-approved study. Chemical shift MR imaging was performed on 1.5- and 3.0-T MR systems from three vendors. Two independent observers measured SII values in five lumbar segments. Inter- and intraobserver agreement was assessed using the interclass correlation coefficients (ICCs). Differences of mean SII values between different field strengths and MR vendors as well as flip angles were compared by using repeated-measures analysis of variance. Differences of mean SII values between different flip angles were also compared by using paired-sample t test. RESULTS Inter- and intra-observer correlation coefficients showed good agreement (all ICC > 0.75) when measuring SII values at different MR systems (ICCs ranging from 0.896 to 0.983) and flip angles (ICCs ranging from 0.824 to 0.983). There were no significant differences in mean SII values measured by different MR vendors with different field strengths (all p > 0.05 ranging from 0.337 to 0.824). The differences in the mean SII between the four different flip angles were statistically significant (all p < 0.05 ranging from < 0.001 to 0.004) except the group of flip angle 50° versus 70° (p = 0.116). CONCLUSION The SII measurement using chemical shift MR imaging may be comparable between different MR systems. Also high flip angles showed better stability to quantitate lumbar fat content.
منابع مشابه
Assessment of the Characteristics of MRI Coils in Terms of RF Non-Homogeneity Using Routine Spin Echo Sequences
Introduction: One of the major causes of image non-uniformity in MRI is due to the existence of non-homogeneity in RF receive and transmit. This can be the most effective source of error in quantitative studies in MRI imaging. Part of this non-homogeneity demonstrates the characteristics of RF coil and part of it is due to the interaction of RF field with the material being imaged...
متن کاملNon-uniformity of Clinical Head, Head and Neck, and Body Coils in Magnetic Resonance Imaging (MRI)
Introduction Signal intensity uniformity in a magnetic resonance (MR) image indicates how well the MR imaging (MRI) system represents an object. One of the major sources of image non-uniformity in high-field MRI scanners is inhomogeneity of radio-frequency coil. The aim of this study was to investigate non-uniformity in head, head and neck, and body coils and compare the obtained results to det...
متن کاملبررسی شدت نسبی سیگنال تصاویر Magnetic Resonance در بافت رترودیسکال و عضله پتریگوئید خارجی و رابطه آن با یافتههای Magnetic Resonance Imaging
Background and Aims: Disc displacement is the most common temporomandibular joint disorder and magnetic resonance imaging (MRI) is the gold standard in its diagnosis. This disorder can lead to changes in signal intensity of magnetic resonance (MR). The purpose of this study was evaluation of correlation between relative signal intensity of MR images of retrodiscal tissue, superior and inferior ...
متن کاملEffect of Physiological noise on Thoraco-Lumbar spinal cord fMRI in 3T Magnetic field
Introduction: Functional MRI methods have been used to study sensorimotor processing in the brain and the Spinal cord. However, these techniques confront unwanted contributions to the measured signal from physiological fluctuations. For the spinal cord imaging, most of the challenges are consequences of cardiac and respiratory movement artifacts that are considered as signifi...
متن کاملThe modeling of induced current density in eyes from static magnetic fields produce by MR scanner
Introduction: Staff and patient Movement in static magnetic field MRI scanner induces current density in the human tissues, so cause biologic effects in people. The aim of this study was the Modelling of current density induced by moving individual with different velocities in static magnetic field of magnetic resonance imaging. Materials and Methods: current ...