Whether noninvasive optimization of AV and VV delays improves the response to cardiac resynchronization therapy.

نویسندگان

  • Bożena Urbanek
  • Michał Chudzik
  • Artur Klimczak
  • Marcin Rosiak
  • Joanna Lewek
  • Jerzy Krzysztof Wranicz
چکیده

BACKGROUND Device optimization is not routinely performed in patients who underwent cardiac resynchronization therapy (CRT) device implantation. Noninvasive optimization of CRT devices by measurement of cardiac output (CO) can be used as a simple method to assess ventricular systolic performance. The aim of this study was to assess whether optimization of atrioventricular (AV) and interventricular (VV) delay can improve hemodynamic response to CRT and whether this optimization should be performed for each patient individually. METHODS Twenty patients with advanced heart failure New York Heart Association (NYHA) class III/IV, left ventricular ejection fraction ≤ 35% and left bundle branch block (QRS ≥ 120 ms) in sinus rhythm were evaluated from 24 h to 48 h after implantation of a CRT device by means of impedance cardiography (ICG). CO was first measured at each patient's intrinsic rhythm. Patients then underwent adjustments of AV and VV delay from 80 ms to 140 ms and from -60 ms to +60 ms, respectively in 20 ms increment steps and CO at each setting was measured by ICG. Both AV and VV delays were programmed according to the greatest improvement in CO compared to intrinsic rhythm. RESULTS There was a statistically signifi cant increase in CO measured at the intrinsic rhythm compared to different AV delay by mean of 21% (3.8 ± 1.0 vs. 4.6 ± 0.1 L/min, p < 0.05). Optimal AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing caused additional increased CO from intrinsic rhythm by mean of 32.6% (3.8 ± 1.0 vs. 5.04 ± ± 1.0 L/min, p < 0.05). Optimal AV/VV setting delays also resulted in improved hemodynamic responses compared to VV factory setting delay. CONCLUSIONS Both AV and VV delay optimization should be performed in clinical practice. Optimal AV delay improved outcome. However, combination of optimized AV/VV delays provided the best hemodynamic response. Optimized AV/VV delays with left ventricle-preexcitation or simultaneous biventricular pacing increased hemodynamic output compared to intrinsic rhythm and VV factory setting delay.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Cardiac resynchronization therapy during rest and exercise: comparison of two optimization methods

AIMS Optimal exercise programming of cardiac resynchronization therapy (CRT) devices is unknown. We aimed to: (i) investigate variations in optimal atrioventricular (AV) and interventricular (VV) delays from rest to exercise, assessed by both echocardiography and an automated intracardiac electrogram (IEGM) method; (ii) evaluate the acute haemodynamic impact of CRT optimization performed during...

متن کامل

Measurement precision in the optimization of cardiac resynchronization therapy.

BACKGROUND Cardiac resynchronization therapy improves morbidity and mortality in appropriately selected patients. Whether atrioventricular (AV) and interventricular (VV) pacing interval optimization confers further clinical improvement remains unclear. A variety of techniques are used to estimate optimum AV/VV intervals; however, the precision of their estimates and the ramifications of an impr...

متن کامل

Acute Beneficial Hemodynamic Effects of a Novel 3D-Echocardiographic Optimization Protocol in Cardiac Resynchronization Therapy

BACKGROUND Post-implantation therapies to optimize cardiac resynchronization therapy (CRT) focus on adjustments of the atrio-ventricular (AV) delay and ventricular-to-ventricular (VV) interval. However, there is little consensus on how to achieve best resynchronization with these parameters. The aim of this study was to examine a novel combination of doppler echocardiography (DE) and three-dime...

متن کامل

Impact of VV optimization in relation to left ventricular lead position: an acute haemodynamic study.

AIMS Left ventricular (LV) lead placement to the most delayed segment offers the greatest potential benefit to cardiac resynchronization therapy (CRT). We assessed the impact of interventricular (VV) optimization on acute changes in cardiac output (CO) in patients with and without LV pacing of the most delayed segment. METHODS AND RESULTS In 124 patients, the most delayed segment was defined ...

متن کامل

Echocardiographic optimization of the atrioventricular and interventricular intervals during cardiac resynchronization.

An optimized atrioventricular (AV) interval can maximize the benefits of cardiac resynchronization therapy (CRT). If programmed poorly, it may curtail beneficial effects of CRT. AV optimization will not convert non-responder to responder, but may convert under-responder to improved status. There are many echocardiographic techniques for AV optimization but there is no universally accepted gold ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Cardiology journal

دوره 20 4  شماره 

صفحات  -

تاریخ انتشار 2013