A laboratory comparison of four positive pressure ventilators used in the home.
نویسندگان
چکیده
Four brands of ventilator used for noninvasive positive pressure ventilation in the home were assessed to determine their performance on a patient simulator. We tested the tidal volume (VT) preset Companion 2801 (Puritan Bennett), minute volume preset Monnal D (Taema), and two pressure preset ventilators, the Nippy (Friday Medical) and the Bilevel Positive Airway Pressure (BiPAP) (Respironics). A patient simulator was employed to investigate the relationships between VT, peak airway pressure (PAP) and mean airway pressure (MAP), the responses to an additional leak in the circuit and patient effort of a variable duration, which was modelled using a negative pressure pump. For equivalent VTs, >300 mL, the Monnal D and Nippy generated a PAP up to 100% greater than the Companion 2801 and BiPAP. When an additional leak was introduced to the circuit, the VT of the Companion 2801 and Monnal D fell by >50%, while the Nippy and the BiPAP responded by increasing flow and maintained VT close to the level achieved with no leak. When the ventilators were triggered by a simulated patient effort of 0.25 s duration with a frequency 33% greater than that of the ventilator, the minute volume increased by 41% for the Companion 2801, by 18% for the Monnal D (no change expected), and by 17% for the Nippy (less than expected), and fell by 7% for the BiPAP due to irregular triggering. When patient effort was prolonged to 1.5 s, a similar length to the inspiratory time of the ventilators, there was no further change in the minute volume of the Companion 2801 and Monnal D, while that of the Nippy and of the BiPAP increased by 38 and 71%, respectively, compared to baseline. These results show that distinct brands of ventilator respond to changes in the patient and patient circuit in different ways, which are not always predictable from a simple description of their operating principles. This should be borne in mind when choosing a positive pressure ventilator for noninvasive ventilation.
منابع مشابه
Performance of ventilators for noninvasive positive-pressure ventilation in children.
The aim of the present study was to evaluate the performance characteristics of all the ventilators proposed for home noninvasive positive-pressure ventilation in children in France. The ventilators (one volume-targeted, 12 pressure-targeted and four dual) were evaluated on a bench which simulated six different paediatric ventilatory patterns. For each ventilator, the quality of the inspiratory...
متن کاملPerformance of transport ventilator with patient-triggered ventilation.
OBJECTIVES Transport ventilators with inspiratory triggering functions and pressure support-control modes have recently become commercially available. We evaluated these ventilators in comparison with a standard ICU ventilator. STUDY DESIGN Laboratory study with a mechanical lung model. METHODS We compared the performance of four transport ventilators (model 740, Mallinckrodt, Pleasanton, C...
متن کاملAn overnight comparison of two ventilators used in the treatment of chronic respiratory failure.
Differences between bilevel ventilators used for noninvasive intermittent positive pressure ventilation (NIPPV) have been demonstrated during bench testing. However, there are no clinical studies comparing these machines. The authors have previously shown that the Quantum pressure support ventilator and Sullivan variable positive airway pressure II ST differ in performance during bench testing....
متن کاملComparison of pressure and volume preset nasal ventilator systems in stable chronic respiratory failure.
Nasal intermittent positive pressure ventilation (NIPPV) has been widely used in the treatment of chronic respiratory disease. Ventilators may be volume or pressure preset; each type has theoretical advantages, but to date there has been no formal comparison. We wanted to assess the efficacy on blood gas changes that may be achieved and overall acceptability of four nasal ventilators (two press...
متن کاملDifferent characteristics of ventilator application between tracheostomy- and noninvasive positive pressure ventilation patients with amyotrophic lateral sclerosis
The aim of the study was to investigate the appropriate home ventilator settings for patients with amyotrophic lateral sclerosis (ALS).In total, 71 patients with ALS, who had received either a noninvasive positive pressure ventilation (NIPPV) or tracheostomy positive pressure ventilation (TPPV), were included. Accordingly, patients were divided into 2 groups (the TPPV and NIPPV groups). We retr...