Comparison of the point-of-care blood gas analyzer versus the laboratory auto-analyzer for the measurement of electrolytes

نویسندگان

  • Anunaya Jain
  • Imron Subhan
  • Mahesh Joshi
چکیده

BACKGROUND Electrolyte values are measured both by arterial blood gas (ABG) analyzers and central laboratory auto-analyzers (AA), but a significant time gap exists between the availability of both these results, with the ABG giving faster results than the AA. The authors hypothesized that there is no difference between the results obtained after measurement of electrolytes by the blood gas and auto-analyzers. METHODS After approval by the ethics committee, an observational cohort study was conducted in which 200 paired venous and arterial samples from patients admitted to the Medical Intensive Care Unit (ICU) of Apollo Hospital, Hyderabad, India, were analyzed for electrolytes on the ABG machine and the AA. Analyses were done on the ABL555 blood gas analyzer and the Dade Dimension RxL Max, both located in the central laboratory. Statistical analyses were performed using paired Student's t test. RESULTS A total of 200 paired samples were analyzed. The mean ABG sodium value was 131.28 (SD 7.33), and the mean AA sodium value was 136.45 (SD 6.50) (p < 0.001). The mean ABG potassium value was 3.74 (SD 1.92), and the mean AA potassium value was 3.896 (SD 1.848) (p = 0.2679). CONCLUSION Based on the above analysis, the authors found no significant difference between the potassium values measured by the blood gas machine and the auto-analyzer. However, the difference between the measured sodium was found to be significant. We therefore conclude that critical decisions can be made by trusting the potassium values obtained from the arterial blood gas analysis.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of point-of-care versus central laboratory measurement of electrolyte concentrations on calculations of the anion gap and the strong ion difference.

BACKGROUND Clinicians calculate the anion gap (AG) and the strong ion difference (SID) to make acid-base diagnoses. The technology used is assumed to have limited impact. The authors hypothesized that different measurement technologies markedly affect AG and SID values. METHODS SID and AG were calculated using values from the point-of-care blood gas and electrolyte analyzer and the central ho...

متن کامل

Comparison of blood gas and electrolyte test results from the Gem-Premier and the ABL-70 versus a conventional laboratory analyzer.

Blood gas analyzers serve a critical role in providing information that reflects patient homeostasis. This study was undertaken to evaluate the accuracy, reliability, consistency, and bias of the Radiometer ABL-70 point of care blood gas analyzer. Thirty samples were gathered for analysis of pH, pCO2, pO2, sodium, potassium, sodium bicarbonate, and base excess. Twenty-nine samples were gathered...

متن کامل

A Comparative Study on Instrumental Precision of Refrigerated and Non-Refrigerated Auto-Analyzers in Order to Improve Quality Assurance in Biochemistry Laboratory

Background and Objective: Quality control is one of the most important components in order to improve quality assurance in laboratories during analytical steps. For this purpose, coefficient of variation plays an important role. Due to the fast improvement in technology, application of inferential statistics for the comparisons ...

متن کامل

Evaluation of the i-STAT Portable Clinical Analyzer for point-of-care blood testing in the intensive care units of a university children's hospital.

We evaluated the analytical performance of the i-STAT Portable Clinical Analyzer (PCA), a point-of-care testing system consisting of a hand-held analyzer and single-use cartridges that measure different panels of electrolytes, metabolites, blood gases, and hematocrit in 65-100 microl of blood. Our objective was to determine whether PCA measurements at the bedside of patients in the neonatal and...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 2  شماره 

صفحات  -

تاریخ انتشار 2009