Evaluation of an alternative radiochemical purity testing method for technetium-99m sestamibi.

نویسندگان

  • A L Luebke
  • D M Wilary
  • D W Mahoney
  • J C Hung
چکیده

OBJECTIVE Our study evaluated the accuracy and reliability of 3 radiochemical purity (RCP) measurement methods of 99mTc-sestamibi. A regular-sized (1.0 cm x 9.0 cm) Whatman 31 ET Chr paper strip (regular 31 ET) also was included in our evaluation because of its ease in handling. METHODS The miniaturized and regular 31 ET methods were compared with the standard RCP testing method (aluminum oxide-coated plastic thin-layer chromatography [TLC] plate, with > or = 95% ethanol as the developing solvent). A total of 30 experimental runs were performed in triplicate (n = 90) over an RCP range of 82%-98%. The 99mTc-sestamibi preparations were reconstituted purposely to ensure that 50% of the tested samples had RCP values below the 90% limit. RESULTS The evaluated RCP ranges were 89.9% +/- 6.3%, 91.0% +/- 3.8%, and 91.4% +/- 4.3% for the TLC, miniature 31 ET, and regular 31 ET methods, respectively (n = 30 each). A strong correlation was found between the TLC and miniature 31 ET methods (r = 0.92), as well as between the TLC and regular 31 ET methods (r = 0.94). Both alternative methods tended to overestimate RCP value as determined by the TLC method, especially in an RCP range below 95%. This resulted in a false-positive rate of 27% for the miniature 31 ET method and 33% for the regular 31 ET method. The test/retest reliability was 99% for both the TLC and regular 31 ET methods, and 91% for the TLC and miniature 31 ET methods. CONCLUSION The miniature and regular 31 ET methods produced a high false-positive rate, which makes them unacceptable for the determination of RCP value of 99mTc-sestamibi.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

NUCLEAR PHARMACY Rapid Quality Control of Technetium-99m-2-Methoxy Isobutyl Isonitrile (Technetium-99m-Sestamibi)

Objective: Technetium-99m-sestamibi (-1"'Tc-MIBI) is a radiopharmaceutical imaging agent useful for assessing myocardial perfusion. We present a simple, rapid and accurate SepPak® method to confirm the radiochemical purity of "'Tc-MIBI. Methods: Technetium-99m-MIBI preparations of high (>95%), intermediate (83-90%) and low (25-43%) radiochemical purity were evaluated by the recommended thinlaye...

متن کامل

Properties, Uses, Radiochemical Purity and Calibration of Tc-99m.

In the first part of this two-part article, the characteristics, potential uses, radiochemical purity and method of evaluating the activity of Tc99m will be discussed. In the second part of the article, the dosimetry of this radionuclide will be presented for various routes of administration as well as for various chem ical configurations, such as the pertechnetate ion, the colloid and the prot...

متن کامل

Technetium-99m radiopharmaceutical preparation problems: 12 years of experience.

OBJECTIVE Chemical reactions involved in preparing 99mTc radiopharmaceuticals occasionally result in products of substandard purity. A retrospective examination of preparation problems that occurred in the author's institution was conducted to better define the incidence, recognize patterns and identify causes of substandard 99mTc radiopharmaceutical products. METHODS All 99mTc radiopharmaceu...

متن کامل

A new two-strip TLC method for the quality control of technetium-99m mercaptoacetyl-triglycine (99mTc-MAG3)

Background 99mTc-mercaptoacetyl-triglycine (99mTc-MAG3) has been used for dynamic renal imaging since about 30 years. Free pertechnetate (99mTcO4), colloidal 99mTc ((99mTcO2)n), 99mTc-tartrate (precursor), precomplexes (99mTc-(MAG3)x) and lipophilic 99mTc-MAG2 are the main radiochemical impurities that may occur in the preparation. The total amount of these impurities has to be identified befor...

متن کامل

Synthesis, optimization and biological evaluation of 99mTc-digoxin as possible cardiac imaging agent.

Heart imaging radiopharmaceuticals could improve the diagnostic value of routine heart scanning for detecting cardiac disorders. The aim of the study was to prepare high radiochemical purity 99mTc-Digoxin in a yield of about 98%. The optimal conditions for labelling were as follows: 100μg of Digoxin, 2μg of SnCl2•2H2O, room temperature (25±1°C), reaction retention time of ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of nuclear medicine technology

دوره 28 4  شماره 

صفحات  -

تاریخ انتشار 2000