Influence of Ionizing Radiation on Two Generations of Cochlear Implants

نویسندگان

  • Nicolas Guevara
  • Anaïs Gérard
  • Jeanne Dupré
  • Delphine Goursonnet
  • Michel Hoen
  • Dan Gnansia
  • Gaëlle Angellier
  • Juliette Thariat
چکیده

The purpose of the present study was to test the behavior of two different generations of cochlear implant systems subjected to a clinical radiotherapy scheme and to determine the maximal acceptable cumulative radiation levels at which the devices show out-of-specification behaviors. Using stereotactic irradiation (Cyberknife, 6 MV photon beam), three Digisonic SP and three Neuro devices were submitted to 5 Gy doses that cumulated to 60 Gy (12 sessions) and 80 Gy (16 sessions), respectively. A follow-up series of irradiation was then applied, in which Digisonic SP devices received two additional fractions of 50 Gy each, cumulating to 160 Gy, and Neuro devices three additional fractions of 20, 40, and 150 Gy, cumulating to 290 Gy. Output current values were monitored during the treatment. At clinical doses, with 60 or 80 Gy cumulative radiation exposure, no single measurement showed more than 10% divergence from the reference measure. The cochlear implants tested in this study showed high resistance to clinically relevant cumulative radiation doses and showed no out-of-bounds behavior up to cumulative doses of 140 or 160 Gy. These observations suggest that cochlear implant users can undergo radiotherapy up to cumulative doses well above those currently used in clinical situations without risk of failure.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of Auditory Perception in Cochlear Implanted Children with and without Additional Disabilities

Background: The number of children with cochlear implants who have other difficulties such as attention deficiency and cerebral palsy has increased dramatically. Despite the need for information on the results of cochlear implantation in this group, the available literature is extremely limited. We, therefore, sought to compare the levels of auditory perception in children with cochlear implant...

متن کامل

Comparison of the high-frequency morpho-syntactic structures of cochlear implant children and children with normal hearing aged 4-6 years

Introduction: Children with cochlear implants experience problems at all language domains, and have more problems in morpho-syntactic skills than others domains. Considering the importance of morphology and syntax in developing of communication skills of children, this study compared the use of high-frequency morpho-syntactic structures among 4-6 years old children with cochlear implants and ty...

متن کامل

Study and Comparison of Formant Characteristics of Persian Vowels in 4-7-year-old Children Using Cochlear Implants and Those Using Hearing Aids

Background and Objective: One of the most important physical properties of vowels is their formant structure. One of the most obvious speech errors in hearing-impaired children is vowel errors. The present study aimed to determine and compare the formant structure of Persian vowels in deaf and cochlear implant children in the age range of 4-7 years. Materials and Methods: This descriptive-anal...

متن کامل

Group Positive Psychotherapy Effect on Increasing the Happiness of Mothers of Children with Cochlear Implant

Objective: The present study aimed to investigate the effects of group positive psychotherapy on increasing the happiness of mothers of children with cochlear implants.  Methods: This study was a quasi-experimental research with pretest-posttest, a control group, and a 75-day follow-up design. A total of 20 mothers of children with cochlear implants treated in Baqiyatallah Hospital of Teh...

متن کامل

Cobalt Poisoning: A Comprehensive Review of the Literature

Cobalt is a chemical element that has been used for centuries recreationally as pigments for colouring, giving the bluish hue to glass. Cobalt Poisoning can be manifested by a wide array of systemic effects. Cobalt poisoning can be seen as a result of super alloys in orthopaedic implants, ionizing radiation, occupational exposure and rarely as a foam stabilizer in beer. Here in we conduct a rev...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:

دوره 2015  شماره 

صفحات  -

تاریخ انتشار 2015