Off-pump transapical implantation of artificial neo-chordae to correct mitral regurgitation: the TACT Trial (Transapical Artificial Chordae Tendinae) proof of concept.

نویسندگان

  • Joerg Seeburger
  • Mauro Rinaldi
  • Sten Lyager Nielsen
  • Stefano Salizzoni
  • Ruediger Lange
  • Markus Schoenburg
  • Ottavio Alfieri
  • Michael Andrew Borger
  • Friedrich Wilhelm Mohr
  • Audrius Aidietis
چکیده

OBJECTIVES The goal of this study was to evaluate the safety and performance of the NeoChord DS1000 system (NeoChord, Inc., Minneapolis, Minnesota). BACKGROUND There is an increasing interest in transcatheter mitral valve (MV) treatment. The NeoChord DS 1000 system enables off-pump beating heart transapical MV repair with implantation of artificial neo-chordae. METHODS Patients with severe mitral regurgitation (MR) due to isolated posterior prolapse were included in this TACT (Transapical Artificial Chordae Tendinae) trial. All patients were scheduled for off-pump transapical implantation of neo-chordae. RESULTS Thirty patients at 7 centers were enrolled. Major adverse events included 1 death due to post-cardiotomy syndrome and concomitant sepsis and 1 minor stroke with the patient fully recovered at the 30-day follow-up visit. Additional patients experienced procedural major adverse events related to a reoperation or conversion to standard of care. Acute procedural success (placement of at least 1 neo-chord and reduction of MR from 3+ or 4+ to ≤2+) was achieved in 26 patients (86.7%). In 4 patients neo-chordae were not placed for technical and/or patient-specific reasons. These patients underwent intraoperative (3 patients) or post-operative (1 patient) standard MV repair. At 30 days, 17 patients maintained an MR grade ≤2+. Four patients who developed recurrent MR were successfully treated with open MV repair during 30-day follow-up. Results improved with experience: durable reduction in MR to ≤2+ at 30 days was achieved in 5 (33.3%) of the first 15 patients and 12 (85.7%) of the last 14 patients. CONCLUSIONS Off-pump transapical implantation of artificial chordae to correct MR is technically safe and feasible; however, it yields further potential for improvement of efficacy and durability. (Safety and Performance Study of the NeoChord Device [TACT]; NCT01777815).

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Off pump implantation of artificial chordae to correct mitral regurgitation – early results

Methods This data is part of multi-center, nonrandomized, prospective Transapical Artificial Chordae Tendinae [TACT] trial. 13 patients with severe mitral regurgitation caused by isolated posterior mitral valve leaflet prolapse, were included in the study. The operation was done through left anterolateral thoracotomy on a beating heart. With TEE guidance, the NeoChord DS1000 was introduced into...

متن کامل

Images and Case Reports in Interventional Cardiology Transapical Beating Heart Mitral Valve Repair

Repair of mitral valve (MV) leaflet prolapse using chordal replacement is an established procedure in open heart surgery for the treatment of mitral regurgitation (MR). The technique has proven high reproducibility and excellent longterm outcomes for posterior leaflet prolaspe, with freedom from valve-related reoperation of 95%.1,2 Striving toward a lessinvasive treatment for structural heart d...

متن کامل

Off-pump transapical neo-chordae implantation.

A 61-year-old female patient presented with isolated pos ter ior mi t ra l l ea f l e t (PML) pro lapse due to degenerative mitral valve (MV). Preoperative transthoracic echocardiography confirmed severe mitral regurgitation (MR) and isolated prolapse of the P2 segment. The patient was in New York Heart Association (NYHA) functional class II to III. Her left ventricular function was good with a...

متن کامل

Transapical beating heart mitral valve repair.

Repair of mitral valve (MV) leaflet prolapse using chordal replacement is an established procedure in open heart surgery for the treatment of mitral regurgitation (MR). The technique has proven high reproducibility and excellent longterm outcomes for posterior leaflet prolaspe, with freedom from valve-related reoperation of 95%.1,2 Striving toward a lessinvasive treatment for structural heart d...

متن کامل

Pii: S1010-7940(01)00990-3

Objectives: There are an increasing number of reports concerning mitral valve repair by reconstruction of the chordae tendinae with the use of expanded polytetrafluoroethylene (PTFE) sutures. However, little information is available about the extended application or results of this technique for extended prolapse of bilateral mitral leaflets, which is considered to be difficult to repair. Metho...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 63 9  شماره 

صفحات  -