Novel non-occlusive temporary endoluminal neck protection device to assist in the treatment of aneurysms in a canine model.

نویسندگان

  • Raymond D Turner
  • M Imran Chaudry
  • Aquilla S Turk
چکیده

INTRODUCTION Current models for the endovascular treatment of aneurysms often require the use of adjuvant devices. Balloon-assisted techniques require temporary occlusion of the parent vessel. Intravascular stents require the use of antiplatelet medication. A novel device that achieves temporary neck protection without parent vessel occlusion or antiplatelet medication is examined in a canine model. METHOD 20 sidewall wide-necked aneurysms were created in 10 canines, one in each carotid artery, using a vein graft technique, and allowed to mature. In each canine, one aneurysm was catheterized with a microcatheter while the neck protection device (NPD) was unsheathed across the neck of the aneurysm and the aneurysm was coiled. The second aneurysm in each canine was coiled without an adjunctive device. The NPD was assessed for coil herniation, coil entrapment within the device, deployability and retrievability. Five animals were killed immediately and five were killed at 28 days and the carotid artery was explanted and sent for necropsy to assess for injury to the endothelium. RESULTS All aneurysms were successfully treated with no adverse events. The NPD was well visualized and unsheathed and resheathed successfully in all cases. There was no coil entrapment within the device, and there were no cases of coil herniation. There was no evidence of endothelial injury or dissection. CONCLUSION The device was technically successful in all devices and further development of this device may be of benefit in endovascular treatment of aneurysms.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Design an Equivalent Left Ventricular Assist Device for Medical Equipment Labs

LVAD is a mechanical pump supporting a weak heart function and blood flow. Sometimes, the heart may not recover fast enough to take over the pumping action immediately after surgery, in such patients a temporary support device has been employed to maintain the pumping action until the patient’s own heart recovers. This device can be considered as a temporary alternative before the process of ar...

متن کامل

Novel aneurysm neck reconstruction device: initial experience in an experimental preclinical bifurcation aneurysm model

INTRODUCTION Treatment of wide-necked bifurcation aneurysms often poses procedural and long-term outcome challenges. The initial preclinical experience with the Pulsar Vascular Aneurysm Neck Reconstruction Device (PVANRD) in a canine bifurcation model is described. METHODS Experimental bifurcation vein pouch aneurysms were surgically created in the carotid arteries of eight dogs. Endovascular...

متن کامل

A loose-packing approach for coil embolization of giant intracranial aneurysm.

Endoluminal occlusion of giant intracranial aneurysms with coil embolization is a viable endovascular treatment option alternative to surgical clipping. However, due to the relatively large aneurysm size, the use of embolization coils for giant aneurysms could be great. A loose-packing embolization strategy in which the fundus of the aneurysm is loosely packed while the aneurysm base is tightly...

متن کامل

New generation of Flow Diverter (surpass) for unruptured intracranial aneurysms: a prospective single-center study in 37 patients.

BACKGROUND AND PURPOSE In patients harboring intracranial aneurysms, the major goal in treatment is to prevent bleeding. A new generation of an endoluminal device (Surpass Flow Diverter [Surpass]) was developed to reconstruct parent artery and occlude the aneurysm. We present our clinical and angiographic single-center experience. METHODS Patients with a wide range of complex unruptured aneur...

متن کامل

Treatment of intracranial aneurysms by functional reconstruction of the parent artery: the Budapest experience with the pipeline embolization device.

BACKGROUND AND PURPOSE Aneurysm treatment by intrasaccular packing has been associated with a relatively high rate of recurrence. The use of mesh tubes has recently gained traction as an alternative therapy. This article summarizes the midterm results of using an endoluminal sleeve, the PED, in the treatment of aneurysms. MATERIALS AND METHODS A total of 19 wide-neck aneurysms were treated in...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Journal of neurointerventional surgery

دوره 3 3  شماره 

صفحات  -

تاریخ انتشار 2011