Preclinical evaluation of a new hollow fiber silicone membrane oxygenator for pediatric cardiopulmonary bypass: ex-vivo study.
نویسندگان
چکیده
Based on the results of many experimental models, a hollow fiber silicone membrane oxygenator applicable for long-term extracorporeal membrane oxygenation (ECMO) was developed. For further high performance and antithrombogenicity, this preclinical model was modified, and a new improved oxygenator was successfully developed. In addition to ECMO application, the superior biocompatibility of silicone must be advantageous for pediatric cardiopulmonary bypass (CPB). An ex vivo short-term durability test for pediatric CPB was performed using a healthy miniature calf for six hours. Venous blood was drained from the left jugular vein of a calf, passed through the oxygenator and infused into the left carotid artery using a Gyro C1E3 centrifugal pump. For six hours, the O2 and CO2 gas transfer rates were maintained around 90 and 80 ml/min at a blood flow rate of 2 L/min and V/Q=3, respectively. The plasma free hemoglobin was maintained around 5 mg/dl. These data suggest that this newly improved oxygenator has superior efficiency, less blood trauma, and may be suitable for not only long-term ECMO but also pediatric CPB usage.
منابع مشابه
Gas transfer performance of a hollow fiber silicone membrane oxygenator: ex vivo study.
Based on the results of in vitro studies of many experimental models, a silicone hollow fiber membrane oxygenator for pediatric cardiopulmonary bypass (CPB) and extracorporeal membrane oxygenation (ECMO) was developed using an ultrathin silicone hollow fiber with a 300 microm outer diameter and a wall thickness of 50 microm. In this study, we evaluated the gas transfer performance of this oxyge...
متن کاملDevelopment of a new hollow fiber silicone membrane oxygenator for ECMO: the recent progress.
Throughout the last 50 years, many improvements have been made for a more effective oxygenator. A large plate type membrane oxygenator, used by Clowes, and a coil type, used by Kolff, has evolved into the small hollow fiber oxygenator. The complex bubble oxygenator, or rotating disk oxygenator, has become a small disposable bubble oxygenator. The currently available oxygenators are extremely sm...
متن کاملFeasibility of a new hollow fiber silicone membrane oxygenator for long-term ECMO application.
Currently in United States, there are no clinically-applicable hollow fiber extracorporeal membrane oxygenation (ECMO) oxygenators available. Therefore, our laboratory is in the process of developing a silicone hollow fiber membrane oxygenator for long-term ECMO usage. This oxygenator incorporates an ultrathin silicone hollow fiber. At this time, a specially-modified blood flow distributor (one...
متن کاملThe effects of cardiopulmonary bypass with hollow fiber membrane oxygenator on blood clotting measured by thromboelastography.
In cardiac surgical patients we investigated the effects of cardiopulmonary bypass (CPB) with a hollow fiber membrane oxygenator on blood clotting measured by thromboelastography (TEG). We found only a minimal change in the strength of blood clot described either by the TEG parameter MA (maximum amplitude) or by the shear modulus G calculated from MA. After CPB there was also a significant tend...
متن کاملMembrane Oxygenator with a Built-In Hemoconcentrator
In order to facilitate the handling of cardiopulmonary bypass (CPB) and simplify the circuit, we have developed a new membrane oxyganator with a hemofiltration function. The hollow fiber units for gas exchange and hemofiltration were combined in concentric circles in a cylindrical housing. The total priming volume was 190 ml. Because we used a silicon-coated hollow fiber membrane for gas exchan...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- Annals of thoracic and cardiovascular surgery : official journal of the Association of Thoracic and Cardiovascular Surgeons of Asia
دوره 8 1 شماره
صفحات -
تاریخ انتشار 2002