Real-time fluorescence lifetime actuation for cell sorting using a CMOS SPAD silicon photomultiplier.

نویسندگان

  • Francescopaolo Mattioli Della Rocca
  • Jakub Nedbal
  • David Tyndall
  • Nikola Krstajić
  • David Day-Uei Li
  • Simon M Ameer-Beg
  • Robert K Henderson
چکیده

Time-correlated single photon counting (TCSPC) is a fundamental fluorescence lifetime measurement technique offering high signal to noise ratio (SNR). However, its requirement for complex software algorithms for histogram processing restricts throughput in flow cytometers and prevents on-the-fly sorting of cells. We present a single-point digital silicon photomultiplier (SiPM) detector accomplishing real-time fluorescence lifetime-activated actuation targeting cell sorting applications in flow cytometry. The sensor also achieves burst-integrated fluorescence lifetime (BIFL) detection by TCSPC. The SiPM is a single-chip complementary metal-oxide-semiconductor (CMOS) sensor employing a 32×32 single-photon avalanche diode (SPAD) array and eight pairs of time-interleaved time to digital converters (TI-TDCs) with a 50 ps minimum timing resolution. The sensor's pile-up resistant embedded center of mass method (CMM) processor accomplishes low-latency measurement and thresholding of fluorescence lifetime. A digital control signal is generated with a 16.6 μs latency for cell sorter actuation allowing a maximum cell throughput of 60,000 cells per second and an error rate of 0.6%.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

The Digital Silicon Photomultiplier – Principle of Operation and Intrinsic Detector Performance

We developed a fully digital implementation of the Silicon Photomultiplier. The sensor is based on a single photon avalanche photodiode (SPAD) integrated in a standard CMOS process. Photons are detected directly by sensing the voltage at the SPAD anode using a dedicated cell electronics block next to each diode. This block also contains active quenching and recharge circuits as well as a one bi...

متن کامل

Time-Domain Fluorescence Lifetime Imaging Techniques Suitable for Solid-State Imaging Sensor Arrays

We have successfully demonstrated video-rate CMOS single-photon avalanche diode (SPAD)-based cameras for fluorescence lifetime imaging microscopy (FLIM) by applying innovative FLIM algorithms. We also review and compare several time-domain techniques and solid-state FLIM systems, and adapt the proposed algorithms for massive CMOS SPAD-based arrays and hardware implementations. The theoretical e...

متن کامل

Fluorescence lifetime biosensing with DNA microarrays and a CMOS-SPAD imager

Fluorescence lifetime of dye molecules is a sensitive reporter on local microenvironment which is generally independent of fluorophores concentration and can be used as a means of discrimination between molecules with spectrally overlapping emission. It is therefore a potentially powerful multiplexed detection modality in biosensing but requires extremely low light level operation typical of bi...

متن کامل

Real-time fluorescence lifetime imaging system with a 32 x 32 0.13microm CMOS low dark-count single-photon avalanche diode array.

A compact real-time fluorescence lifetime imaging microscopy (FLIM) system based on an array of low dark count 0.13microm CMOS single-photon avalanche diodes (SPADs) is demonstrated. Fast background-insensitive fluorescence lifetime determination is achieved by use of a recently proposed algorithm called 'Integration for Extraction Method' (IEM) [J. Opt. Soc. Am. A 25, 1190 (2008)]. Here, IEM i...

متن کامل

Time-Correlated Fluorescence Microscopy Using a Room Temperature Solid-State Single Photon Sensor

In this paper we present a two-photon fluorescence lifetime imaging microscopy (FLIM) system based on a two-dimensional single photon avalanche diode (SPAD) array fabricated in CMOS [1] [2]. To the best of our knowledge, this is the first demonstration of two-photon FLIM based on a CMOS detector. The sensor reported here exhibits picosecond accuracy, operates at room temperature, and requires n...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • Optics letters

دوره 41 4  شماره 

صفحات  -

تاریخ انتشار 2016