BACKGROUND Despite technical advancements, implantation of coronary sinus (CS) leads may be challenging, and dislocation remains a relevant clinical problem. OBJECTIVE The aim of this study was to investigate the effectiveness, safety, and long-term outcome of stent implantation to anchor the lead to the wall of the CS side branch. METHODS Stenting of a CS side branch was performed in 312 patients. The procedure was performed because of postoperative lead dislocation in 16 patients and because of an intraoperative unstable lead position or phrenic nerve stimulation in 296 cases. A bare metal coronary stent was introduced over a second guide wire in the same CS sheath. The stent was deposited 5-35 mm proximal to the most proximal electrode. Mechanical damage of the CS side branch or pericardial effusion was not observed owing to stenting. RESULTS During follow-up (median 28.4, interquartile range 15-37, maximum 70 months), a clinically important increase in the left ventricular pacing threshold was found in four cases and reoperation was necessary in only two patients (0.6%). Phrenic nerve stimulation was observed in 18 instances, and repositioning with an ablation catheter was performed in seven cases. Impedance measurements did not suggest lead insulation failure. Three stented leads were extracted without complication after 3-49 months owing to infection, while four leads were extracted easily during heart transplantation after 7-27 months. CONCLUSION Stent implantation to stabilize CS lead position seems to be an effective and safe procedure in prevention and treatment of CS lead dislocation in selected cases.
منابع مشابه
First-in-Man Coronary Sinus Lead Stabilization Using a Bioresorbable Vascular Scaffold System.
Cardiac resynchronization therapy has become an integral part of treatment in patients presenting with reduced ventricular function (left ventricular ejection fraction <35%), clinically symptomatic dyspnea (New York Heart Association II–IV), and complete left bundle branch block. Currently, the standard approach of left ven-tricular lead placement is transvenously via the coronary sinus (CS). A...
متن کاملAcute and Long Term Outcomes of Coronary Intervention in Unprotected Left Main Lesions
Introduction : Significant left main coronary artery stenosis jeopardizes the entire myocardium of the left ventricle and has the worst prognosis of any form of coronary artery disease. Coronary-artery bypass grafting (CABG) has been considered as the standard therapeutic approach for such patients. There are limited data on the safety and effectiveness of percutaneous coronary intervention (PC...
متن کاملImplication of a persistent left superior vena cava in transvenous pacemaker therapy and cardiac hemodynamic monitoring.
An anomalous persistent left superior vena cava may complicate the insertion of transvenous pacemaker electrode catheters as well as the catheterization of the pulmonary artery. It is important to recognize the anomaly because of the ease of confusing the position of the catheters pas+ ing through the coronary sinus into a distal coronary vein rather than the right ventricular apex. If the cond...
متن کاملComment on: Dual LAD with anomalous origin of long LAD from right coronary sinus: A variant of type VI LAD
A 50‐year‐old gentleman was referred to our department for computed tomography (CT) coronary angiography with complaints of pain on the left side of the chest and cold sweats since 4 days. CT coronary angiography, done on a 64‐slice CT (Somatom Sensation‐64, Erlangen, Germany), showed evidence of dual LAD. The short LAD was arising from the left main coronary artery and coursing in the proximal...
متن کاملImpact of stenting technique and bifurcation anatomy on long-term outcomes of PCI for distal unprotected left main coronary disease.
OBJECTIVES We aimed to assess the associations of stenting strategy and bifurcation anatomy with outcomes of percutaneous coronary intervention (PCI) for distal unprotected left main (ULM) coronary disease. BACKGROUND There are limited and conflicting data regarding long-term outcomes associated with stenting strategies for PCI of distal ULM coronary disease. METHODS Patients undergoing non...