Rise in chronic defibrillation energy requirements necessitating implantable defibrillator lead system revision.
نویسندگان
چکیده
The chronic defibrillation energy requirement (DER) is believed to remain clinically stable in patients with defibrillators. Six patients (two with an epicardial and four with a nonthoracotomy system) were identified with a rise in their chronic DER, which eliminated a 10-J safety margin, thus necessitating a defibrillator lead system revision. The mean increase in DER was 14.7 +/- 4 J and was discovered at a mean of 16.0 +/- 18 months (range 2-41) following implantation. Management included placement of a defibrillator with a biphasic waveform, placement of an additional defibrillation electrode, or both. At 2 months following revision of the defibrillation system, a 10-J DER safety margin was present in each patient. In some patients, there is a progressive increase in the chronic DER with elimination of a 10-J safety margin necessitating revision of the defibrillation system. Routine reevaluation of the chronic DER, therefore, is necessary to identify these patients.
منابع مشابه
Long-term evaluation of the ventricular defibrillation energy requirement.
INTRODUCTION Defibrillation energy requirements in patients with nonthoracotomy defibrillators may increase within several months after implantation. However, the stability of the defibrillation energy requirement beyond 1 year has not been reported. The purpose of this study was to characterize the defibrillation energy requirement during 2 years of clinical follow-up. METHODS AND RESULTS Th...
متن کاملIncidence of lead system malfunction detected during implantable defibrillator generator replacement.
Implantable cardioverter-defibrillator (ICD) generator replacement due to a depleted battery is a frequently performed procedure. The frequency with which sensing and defibrillation system failures are identified during device replacement procedures has not been previously described. Therefore, the purpose of this study was to prospectively determine the frequency of lead system malfunction det...
متن کاملEffect of chronic amiodarone therapy on defibrillation energy requirements in humans.
INTRODUCTION The effect of oral amiodarone therapy on defibrillation energy requirements in patients with an implantable defibrillator has not been established. METHODS AND RESULTS Twenty-one consecutive patients with implantable biphasic waveform defibrillators underwent a step-down determination of the defibrillation energy requirement 211 +/- 12 days before and 73 +/- 22 days after initiat...
متن کاملLead systems for internal ventricular fibrillation.
We examined the feasibility of using a standby automatic implantable defibrillator and established thresholds for internal defibrillation. The implantable defibrillator senses ventricular fibrillation and delivers an electrical impulse for defibrillation. Two lead systems for the device have been investigated. System I consists of two 12-cm2 silicone-covered titanium mesh patches attached to th...
متن کاملTransvenous biventricular defibrillation halves energy requirements in patients.
BACKGROUND Defibrillation thresholds (DFT) with standard implantable cardioverter-defibrillator leads in the right ventricle (RV) may be determined by weak shock field intensity in the myocardium of the left ventricle (LV). Adding a shocking electrode in a coronary vein on the middle of the LV free wall, thereby establishing biventricular defibrillation, substantially reduced defibrillation req...