Measuring carboxypeptidase A activity with a centrifugal analyzer: analytical and clinical considerations.
نویسندگان
چکیده
This adaptation of a commercially available kit for automated measurement of carboxypeptidase A (CPA; EC 3.4.17.1) activity in serum with the Cobas Bio centrifugal analyzer extends the linear range to an activity concentration of 82 U/L. Results obtained by the described method correlated closely (r = 0.98) with those by the manual kit method. The reference interval for 150 apparently normal individuals was 0.12-0.91 U/L. Total CVs of the method ranged from 4.0% to 13.1%. Bilirubin and glucose decreased the CPA activity in serum by as much as 98% and 26%, respectively. Substantial CPA activity was found in pancreatic tissue, with little activity in intestinal tissue. CPA activity was not as widely distributed in extra-pancreatic tissues as were amylase and lipase activities. Peak activities of CPA, amylase, and lipase in the sera of patients with acute pancreatitis were significantly correlated (r = 0.45 to 0.78, P less than 0.05-0.01). The optimized diagnostic efficiency of CPA for acute pancreatitis was 0.85 at a cutoff value of 5 U/L. Amylase and lipase exhibited similar optimized efficiencies, and parallel testing did not significantly improve diagnostic accuracy. We conclude that automated analysis for CPA activity, even in the absence of interferences, does not add to the diagnostic information provided by the widely available assays for amylase and lipase activity.
منابع مشابه
Direct, fixed-time kinetic assays for beta-hydroxybutyrate and acetoacetate with a centrifugal analyzer or a computer-backed spectrophotometer.
In the course of studying the control of blood glucose in juvenile onset diabetics, we developed convenient methods for determining beta-hydroxybutyrate and acetoacetate. Here we describe fixed-time, enzymic, reaction-rate procedures for directly measuring these organic acids with a centrifugal analyzer (Rotochem IIA/36) or a computer-backed spectrophotometer (Gilford 102 system). In either cas...
متن کاملBenzethonium chloride method for proteins adapted to centrifugal analysis.
A benzethonium chloride method for measuring protein concentrations in several types of body fluids has been adapted to the Cobas Bio centrifugal analyzer. The ability of this analyzer to add two reagents stepwise, measuring absorbances after each addition, is essential for proper blanking of urine specimens. The method has excellent curve stability and precision and yields results that correla...
متن کاملEvaluation of kinetic light scattering as an approach to the measurement of specific proteins with the centrifugal analyzer. II. Theoretical considerations.
Development of centrifugal analyzers capable of making light-scattering measurements has prompted the investigation of the application of this type of instrumentation to the measurement of specific proteins. Investigation of antigen/antibody complex dimension and the kinetics of the IgG/anti-IgG reaction as followed by light scattering are examined in relation to improved analytical methodology...
متن کاملHomogeneous enzyme immunoassay for theophylline with a centrifugal analyzer.
A rapid homogeneous enzyme immunoassay for theophylline has been adapted to a centrifugal analyzer. The sample volume used was 10 microliter and sera from as many as 28 patients could be measured in 180 seconds. The assay temperature was 30 degrees and the absorbance rates were measured at 340 nm. Within-run and run-to-run precision showed a coefficient of variation of less than 2.6 and 7.5 per...
متن کاملCentrifugal analyzer method for total bilirubin in serum by use of diazotized 2-chloroaniline-5-sulfonic acid.
We describe two centrifugal analyzer methods for measuring total bilirubin in serum. Diazotized 2-chloroaniline-5-sulfonic acid is used in both. In the first procedure, ethylene glycol and methanol are used as an accelerator solvent; results correlate well with those by a Jendrassik and Grof method adapted to the centrifugal analyzer, but there is considerable hemolysis interference. In the sec...