Effect of regulating airway pressure on intrathoracic pressure and vital organ perfusion pressure during cardiopulmonary resuscitation: a non-randomized interventional cross-over study
نویسندگان
چکیده
BACKGROUND The objective of this investigation was to evaluate changes in intrathoracic pressure (Ppl), airway pressure (Paw) and vital organ perfusion pressures during standard and intrathoracic pressure regulation (IPR)-assisted cardiopulmonary resuscitation (CPR). METHODS Multiple CPR interventions were assessed, including newer ones based upon IPR, a therapy that enhances negative intrathoracic pressure after each positive pressure breath. Eight anesthetized pigs underwent 4 min of untreated ventricular fibrillation followed by 2 min each of sequential interventions: (1) conventional standard CPR (STD), (2) automated active compression decompression (ACD) CPR, (3) ACD+ an impedance threshold device (ITD) CPR or (4) ACD+ an intrathoracic pressure regulator (ITPR) CPR, the latter two representing IPR-based CPR therapies. Intrapleural (Ppl), airway (Paw), right atrial, intracranial, and aortic pressures, along with carotid blood flow and end tidal CO2, were measured and compared during each CPR intervention. RESULTS The lowest mean and decompression phase Ppl were observed with IPR-based therapies [Ppl mean (mean ± SE): STD (0.8 ± 1.1 mmHg); ACD (-1.6 ± 1.6); ACD-ITD (-3.7 ± 1.5, p < 0.05 vs. both STD and ACD); ACD-ITPR (-7.0 ± 1.9, p < 0.05 vs. both STD and ACD)] [Ppl decompression (mean ± SE): STD (-6.3 ± 2.2); ACD (-13.0 ± 3.8); ACD-ITD -16.9 ± 3.6, p < 0.05 vs. both STD and ACD); ACD-ITPR -18.7 ± 3.5, p < 0.05 vs. both STD and ACD)]. Interventions with the lower mean or decompression phase Ppl also demonstrated lower Paw and were associated with higher vital organ perfusion pressures. CONCLUSIONS IPR-based CPR methods, specifically ACD-ITPR, yielded the most pronounced reduction in both Ppl and Paw and resulted in the most favorable augmentation of hemodynamics during CPR.
منابع مشابه
Evaluation of a prototypic inspiratory impedance threshold valve designed to enhance the efficiency of cardiopulmonary resuscitation.
OBJECTIVE Assess a prototype inspiratory impedance threshold valve (ITV) designed to enhance vital organ circulation during standard and active compression/decompression cardiopulmonary resuscitation (CPR). BACKGROUND The ITV attaches to commonly used airway assist devices and decreases intrathoracic pressure during the decompression (chest recoil) phase of CPR by creating a vacuum within the...
متن کاملA dose-response curve for the negative bias pressure of an intrathoracic pressure regulator during CPR.
An intrathoracic pressure regulator (ITPR) is a device that can be added to the external end of a tracheal tube to create controlled negative airway pressure between positive pressure ventilations. The resulting downward bias of the airway pressure baseline promotes increased venous return and enhanced circulation during CPR and also during hypovolemic shock. In the present study, we exercised ...
متن کاملUse of an inspiratory impedance threshold device on a facemask and endotracheal tube to reduce intrathoracic pressures during the decompression phase of active compression-decompression cardiopulmonary resuscitation.
INTRODUCTION Use of an inspiratory impedance threshold device (ITD) significantly increases coronary perfusion pressures and survival in patients ventilated with an endotracheal tube (ETT) during active compression-decompression cardiopulmonary resuscitation. We tested the hypothesis that the ITD could lower intratracheal pressures when attached to either a facemask or ETT. METHODS An active ...
متن کاملReducing ventilation frequency during cardiopulmonary resuscitation in a porcine model of cardiac arrest.
INTRODUCTION American Heart Association/American College of Cardiology guidelines recommend a compression-to-ventilation ratio (C/V ratio) of 15:2 during cardiopulmonary resuscitation (CPR) for out-of-the-hospital cardiac arrest. Recent data have shown that frequent ventilations are unnecessary and may be harmful during CPR, since each positive-pressure ventilation increases intrathoracic press...
متن کاملIntrathoracic pressure regulator during continuous-chest-compression advanced cardiac resuscitation improves vital organ perfusion pressures in a porcine model of cardiac arrest.
BACKGROUND A novel device, the intrathoracic pressure regulator (ITPR), combines an inspiratory impedance threshold device (ITD) with a vacuum source for the generation of controlled -10 mm Hg vacuum in the trachea during cardiopulmonary resuscitation (CPR) while allowing positive pressure ventilation. Compared with standard (STD) CPR, ITPR-CPR will enhance venous return, systemic arterial pres...