Dwell position inaccuracy in the Varian GammaMed HDR ring applicator
نویسندگان
چکیده
Varian has issued two Product Notification Letters warning of known inaccuracies in dwell positions for their GammaMed HDR ring applicator. This inaccuracy was measured for two sets of applicators. Autoexposed radiographs were taken of the HDR source at different dwell positions and analyzed per Varian recommendations using tools within the BrachyVision treatment planning program. Comparison between programmed and actual dwell positions showed the actual positions shifted distally by an average of 0.34 cm (0.17 cm-0.59 cm) across all positions in all rings. A correction method was developed and tested. During planning, the tip of the ring was extrapolated distally beyond its actual position in the patient image set and a proximal offset of the same distance was applied to the dwell positions. A global shift of 0.3 mm corrected all but the most proximal actual dwell position to within +2 mm of the planned position.
منابع مشابه
A new approach to measure dwell position inaccuracy in HDR ring applicators – quantification and corrective QA
As part of quality assurance (QA) in high dose rate brachytherapy, it is necessary to verify that the source dwell positions correspond to the radiographic markers used in simulation and treatment planning. The procedure is well established for linear tandem applicators. However, with the advent of ring applicators, this has become more critical and challenging. This work describes a new approa...
متن کاملSU-E-T-138: Quantification of Dwell Position Inaccuracy in Varian GammaMed HDR Titanium Ring Applicators.
PURPOSE To quantify the dwell position inaccuracy in Titanium ring applicators and develop a test to be performed quarterly, after source exchange. METHODS All three rings from our Titanium kit (30, 45, 60 deg.) were used for this study. EDR2 film was placed on the Simulator table and a ring was taped to the film, with a solid water slab as buildup. A 1-cm spacing dummy wire was inserted into...
متن کاملCommissioning of Varian ring & tandem HDR applicators: reproducibility and interobserver variability of dwell position offsets
Studies have shown that source dwells within Varian's HDR CT/MR compatible ring applicators can deviate from intended positions by several millimeters. Quantifying this offset is an important part of commissioning. The aims of this study were to: 1) determine the reproducibility of the offset, 2) study the interobserver variation in the offset's measurement, and 3) quantify the dosimetric impac...
متن کاملQuality assurance measurements for high‐dose‐rate brachytherapy without film
The purpose of this study was to develop new and modified tools that allow HDR brachytherapy quality assurance tests to be carried out efficiently without film, video cameras, stopwatches, and mechanical rulers; and to devise methods that use these new tools for daily and quarterly check procedures, which are efficient and provide increased accuracy compared to previous methods. The HDR brachyt...
متن کاملComparison of dose volume parameters evaluated using three forward planning – optimization techniques in cervical cancer brachytherapy involving two applicators
Purpose This study is intended to compare dose-volume parameters evaluated using different forward planning- optimization techniques, involving two applicator systems in intracavitary brachytherapy for cervical cancer. It looks for the best applicator-optimization combination to fulfill recommended dose-volume objectives in different high-dose-rate (HDR) fractionation schedules. Material and ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
دوره 11 شماره
صفحات -
تاریخ انتشار 2010