Infections after cardioverter-defibrillator implantation: observations in 335 patients over 10 years.

نویسندگان

  • H J Trappe
  • P Pfitzner
  • H Klein
  • P Wenzlaff
چکیده

OBJECTIVE To determine the incidence of infection after implantation of a cardioverter-defibrillator and the management of this complication. SUBJECTS 335 consecutive patients who had a cardioverter-defibrillator implanted between January 1984 and December 1993. MAIN OUTCOME MEASURES Incidence of infection within the first month after implantation (early infection) and after the first month (late infection). RESULTS Infections associated with cardioverter-defibrillator devices occurred in 13 patients (3.9%) during a mean follow up of 22 (11) months. All patients had general signs of inflammation, fever (> 37.5 degrees C), and leucocytosis (> 10,000/ml) with or without purulent drainage. Five patients (38%) had infections during the first implantation, whereas eight patients (62%) had infections after replacement of the pulse generator. Early infection was observed in four patients (31%) and late infection in nine (69%). Incidence of infection was higher in patients who underwent epicardial cardioverter-defibrillator implantation (12/207 patients, 5.8%) than in those who received nonthoracotomy lead systems (1/125 patients, 0.8%) (P < 0.05). Infections were caused by staphyloccocus in 10 patients, pseudomonas in two patients, and streptococcus in one patient. The whole device had to be removed in all patients. During a mean follow up of 39 (29) months seven patients died: six of congestive heart failure and one of myocardial reinfarction. CONCLUSIONS Infection, one of the most serious complications after cardioverter-defibrillator implantation, is associated with increased morbidity and mortality. When infection occurs the system must be removed to avoid a fatal outcome.

برای دانلود رایگان متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Implantable cardioverter-defibrillator in a patient with dextrocardia situs inversus

Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...

متن کامل

THE EFFECT OF NURSE-PEER-LED SUPPORT INTERVENTION ON QUALITY OF LIFE IN PATIENTS WITH AN IMPLANTABLE CARDIOVERTER-DEFIBRILLATOR: A RANDOMIZED CLINICAL TRIAL

Background & Aims: Nowadays, the quality of life and how coping and treatment of patients with an implantable cardiac defibrillator has led to concern and attempts to reduce or eliminate the problems of these patients in the people who provide health care services. This study aimed to determine the effect of nurse-peer-led support intervention on quality of life in patients with an implantable ...

متن کامل

Implantable cardioverter defibrillator specific rehabilitation improves health cost outcomes: Findings from the COPE-ICD randomized controlled trial.

OBJECTIVE The Copenhagen Outpatient ProgrammE - implantable cardioverter defibrillator (COPE-ICD) trial included patients with implantable cardioverter defibrillators in a randomized controlled trial of rehabilitation. After 6-12 months significant differences were found in favour of the rehabilitation group for exercise capacity, general and mental health. The aim of this paper is to explore t...

متن کامل

Annual rate of transvenous defibrillation lead defects in implantable cardioverter-defibrillators over a period of >10 years.

BACKGROUND The number of patients with longer follow-up after implantation of an implantable cardioverter-defibrillator is increasing continuously. Defibrillation lead failure is a typical long-term complication. Therefore, the long-term reliability of implantable cardioverter-defibrillator leads has become an increasing concern. The aim of the present study was to assess the annual rate of tra...

متن کامل

Pacemaker lead endocarditis with hiccups (Kalayci)

Background: Lead-related infections that might develop after pacemaker implantation associated with high mortality and morbidity rates are challenging to manage and pose high-cost. Patients with lead-related infections usually present with fever, chills and fatigue and the treatment can be challenging unless the implant system is extracted. Case presentation: A 66-year old male patient who und...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

عنوان ژورنال:
  • British heart journal

دوره 73 1  شماره 

صفحات  -

تاریخ انتشار 1995