Artifact removal procedure distorts multifocal electroretinogram.

نویسندگان

  • M Yoshii
  • K Yanashima
  • S Suzuki
  • S Okisaka
چکیده

PURPOSE To study whether the Artifact Removal procedure available for eliminating artifacts in multifocal electroretinograms (mERG) works correctly or not. METHODS A test response was made using a photo-diode circuit. mERGs were recorded from 3 well-trained normal subjects using the Veris III system, and were then analyzed by the procedure that is included in the Veris Science (Artifact Removal) software program. The stimuli consisted of densely arranged arrays of 103 or 37 hexagonal elements. It took a total of 8 minutes to obtain one mERG record, and 16 sessions were required to complete this record. The first-order as well as the second-order kernel response components were extracted by Veris Science software, and the Artifact Removal procedure was used for both components. RESULTS The Artifact Removal procedure influenced both the test response on the center element as well as the neighboring traces just around the test response. After the repetitions of the Artifact Removal procedure, the shape of the test response changed considerably. Some of the traces of the second-order kernel response components elicited from a normal subject changed irregularly when the Artifact Removal procedure was repeatedly used. The noise increased at the first iteration of the Artifact Removal procedure. CONCLUSION This procedure has been considered useful for eliminating artifact distortion in mERG, but should be carefully checked by well-established testing methods before clinical use.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Visual loss following removal of intraocular silicone oil.

AIM To investigate the cause of visual loss following removal of intraocular silicone oil in patients who underwent vitrectomies for retinal detachment or giant retinal tear. METHODS The clinical records of three cases with visual loss following removal of silicone oil were reviewed. Investigations carried out included fundus fluorescein angiogram, optical coherence tomography, and electrophy...

متن کامل

Multifocal electroretinogram in clinically significant diabetic macular edema.

OBJECTIVE To explore the multifocal electroretinogram in patients with nonproliferative diabetic retinopathy with clinically-significant macular edema. METHODS Forty-one eyes with clinically significant macular edema were tested. The latencies and amplitudes of average responses of 5 eccentric rings from 0 to 26 degrees relative to the fixation point were compared with normal values obtained ...

متن کامل

Electrophysiological examination in uveitis: a review of the literature

PURPOSE Uveitis is the inflammation of the uveal tract, which usually also affects the retina and vitreous humor. The electrophysiological examination is an objective ocular examination that includes the electroretinogram, visual evoked potentials, the electrooculogram, the multifocal electroretinogram, and multifocal visual evoked potentials. Our aim is to review the literature of the use of t...

متن کامل

Multifocal electroretinogram in birdshot chorioretinopathy.

PURPOSE To characterize multifocal electroretinogram parameters in patients with birdshot chorioretinopathy. METHODS Twenty-eight patients with birdshot chorioretinopathy consecutively included from 2006 to 2011 were matched to 27 healthy subjects for age, axial length, and lens status. Multifocal electroretinogram was prospectively evaluated using the Vision Monitor system. RESULTS Birdsho...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Japanese journal of ophthalmology

دوره 44 4  شماره 

صفحات  -

تاریخ انتشار 2000